Order Entry
United States
ContactUsLinkComponent
 
Nitric Oxide Macro Sensor
  77664-474
New Product
undefined
Nitric Oxide Macro Sensor
  77664-474
 :  ISO-NOP
New Product
Restricted Products: To process your orders without delay, please provide the required business documentation to purchase this product.

To order chemicals, medical devices, or other restricted products please provide ID that includes your business name & shipping address via email [email protected] or fax 484.881.5997 referencing your VWR account number. Acceptable forms of ID are:

  • • State issued document with your organization's Federal Tax ID Number
  • • State issued document with your organization's Resale Tax ID Number
  • • City or County issued Business License
  • • State Department of Health Services License
  • • Any other ID issued by the State that includes the business name & address

* ATTN: California Customers may require additional documentation as part of the CA Health & Safety Code. Products that fall under this regulation will be placed on a mandatory 21-day hold after documentation is received. Avantor will not lift restrictions for residential shipping addresses.

 

  • Ø:
    2 mm
  • Description:
    Nitric oxide sensor
  • Cat. No.:
    77664-474

 

 

The ISO-NOP is a popular, robust and high performance sensor encased within a 2 mm diameter disposable stainless steel protective sleeve. The tip of the sleeve is covered with a NO-selective membrane. Replacement membrane sleeves can be purchased separately and require an internal electrolyte.

  • Excellent sensitivity to NO: 2 pA/nM
  • Rapid response time: <5 s
  • Highly sensitive
  • Lowest detection limit: 1 nM
  • Upper detection limit/range: 80 µM (beyond 80 µM, sensor accuracy is not guaranteed)
  • 2 mm NO sensors for NO detection in cell cultures

A simple change in experimental protocol enables the ISO-NOP to be conveniently used for indirect rapid accurate determination of nitrite (NO₂) and nitrate (NO₃) concentration in samples. Using this method a detection limit for NO₂ or NO₃ as low as 1 nM is routinely possible.

Ideal for cell cultures, cell suspensions and many other applications.

Nominal sensitivity (New sensor): ≤2 pA/nM
Interferent (selectivity coefficient): NaNO₂ (10⁻⁶ or better)
Drift: <1 pA/min.
Poise voltage: 865 mV
Typical quiescent baseline current, 25 °C: 3000 pA
Acceptable baseline range: 1000 to 8000 pA
Polarization time: 12+ hours