Gas detection tubes for benzene
Supplier: Dräger
Certificates
About this item
To measure the 'spot concentrations' of carbon dioxide for use with an Accuro® pump. Uses include determination and measurement of concentration peaks, personal exposure in the inhalation area, detection of leaks in pipelines or gas installations, air analysis in sewers, shafts and confined spaces.
- Benzene 5/a and 15/a with prelayer retains toluene and xylene causing a reddish brown discolouration
- If xylene and toluene levels are too high, the entire prelayer up to the indicating layer is descoloured making benzene mesurement impossible
- Petroleum hydrocarbons, alcohols, esters do not affect the indication for these two tubes
One stroke is equivalent to a volume of 100 ±5 ml of gas passing through the tube.
Specifications
- Catalog No:
- 151-0169
- DRAE8103691
- 118-0289
- 118-0317
- DRAE6718801
- Colour change:
- White to brown grey
- Light grey to dark grey to black
- White to brown green
- White to red brown
- White to red brown
- Humidity:
- 1 - 15 mg/l H₂O
- 3 - 15 mg/l H₂O
- ≤ 30 mg/l H₂O
- ≤ 50 mg/l H₂O
- Measurement time:
- 8 minutes
- 8 minutes
- 4 minutes
- 3 minutes
- No. of strokes:
- 20
- 20
- 20 - 2
- 15 - 2
- Standard deviation:
- ±10 - 15 %
- ±10 - 15 %
- ±30 %
- ±30 %
- Standard range:
- Calibrated at 20 °C and 1013 hPa
- Calibrated at 20 °C and 1013 hPa
- Calibrated at 20 °C and 1013 hPa
- Calibrated at 20 °C and 1013 hPa
- Temperature range:
- 0…40 °C
- 0…40 °C
- 0…30 °C
- 0…40 °C
Specifications