About this item
Spectroquant® Prove spectrophotometers are designed for any type of water analysis - wastewater, process water or drinking water and beverages.
Barcode reading for all test kits – Cell tests AND reagent tests - making the use of analytical test kits as easy as never before
Live ID: the Prove plus instruments automatically read the code, detect the method, and select the right settings. Test-specific lot and expiry data are saved with each result
Measurements can include dilution factors for each sample, turbidity correction, simultaneous display of absorption and concentration, setting of own limits and trend analyses
Removable cell holder for easy cleaning
Test kits can be stored at room temperature
The Spectroquant® Prove 100 plus is the best choice for those who routinely use test kits, or only perform VIS measurements.
The Spectroquant® Prove 300 plus is ideal for intense use due to its the long-lasting Xenon lamp and for UV/VIS measurements.
The Spectroquant® Prove 600 plus is designed with high-end UV/VIS optics, and cuvettes sizes of up to 100 mm. It provides excellent resolution and sensitivity for complex kinetics or spectra in addition to test kit measurements.
All Spectroquant® Prove plus instruments provide high quality and great value for money for your daily analyses. They are pre-programmed for over 150 Spectroquant® test kits and additional >150 methods, giving you swift, reliable results for the broadest range of applications.
All Prove plus instruments automatically recognize the cell size (10,20,50 mm) and select the right method – no adapter needed. Prove 600 plus is also compatible with 100 mm cells.
Specifications
- Catalog No:
- 1.73026.0001
- 1.73027.0001
- 1.73028.0001
- Model:
- Prove 100 plus
- Prove 300 plus
- Prove 600 plus
- Measuring modes:
- Concentration, absorbance, transmission, multi wavelengths, spectra and kinetics in absorbance and
transmission mode - Concentration, absorbance, transmission, multi wavelengths, spectra and kinetics in absorbance and
transmission mode - Concentration, absorbance, transmission, multi wavelengths, spectra and kinetics in absorbance and
transmission mode
- Display:
- p-cap glass touch screen
- p-cap glass touch screen
- p-cap glass touch screen
- Absorbance accuracy:
- at 340 - 900 nm
1 absorbance: ±0,005 absorbance
2 absorbance: ±0,005 absorbance
2,5 absorbance: ±0,010 absorbance - at 300 - 900 nm
1 absorbance: ±0,005 absorbance
2 absorbance: ±0,005 absorbance
2,5 absorbance: ±0,008 absorbance - at 230 - 900 nm
1 absorbance: ±0,004 absorbance
2 absorbance: ±0,004 absorbance
2,5 absorbance: ±0,006 absorbance
- Photometric range:
- ±3,0 Abs
- ±3,0 Abs
- ±3,3 Abs
- Wavelength accuracy:
- ±1 nm
- ±1 nm
- ±1 nm
- Wavelength range:
- 320 - 1100 nm
- 190 - 1100 nm
- 190 - 1100 nm
- Wavelength reproducibility:
- ±0,2 nm
- ±0,2 nm
- ±0,1 nm
- Wavelength resolution:
- 1 nm (scan 0,1 nm)
- 1 nm (scan 0,1 nm)
- 1 nm (scan 0,1 nm)
- Cells:
- 16 mm round cells, 10, 20 and 50 mm rectangular cells with automatic recognition
- 16 mm round cells, 10, 20 and 50 mm rectangular cells with automatic recognition
- 16 mm round cells, 10, 20, 50 and 100 mm rectangular cells with automatic recognition
- Communication interface:
- USB: 2 × USB-A (for printer, USB memory devices, keyboard or bar code reader),
1 × USB-mini-B; Ethernet: LAN connection - USB: 2 × USB-A (for printer, USB memory devices, keyboard or bar code reader),
1 × USB-mini-B; Ethernet: LAN connection - USB: 2 × USB-A (for printer, USB memory devices, keyboard or bar code reader),
1 × USB-mini-B; Ethernet: LAN connection
- Detector:
- Silicon photo diode
- -
- -
- Electrical:
- 100 - 230 V, 50 - 60 Hz
- 100 - 230 V, 50 - 60 Hz
- 100 - 230 V, 50 - 60 Hz
- Lamp source, lifetime:
- Tungsten halogen lamp
- Xenon flash lamp
- Xenon flash lamp
- Languages:
- 28
- 28
- 28
- Max. humidity:
- Operation: 20 - 80% rH, storage in ambient relative humidity conditions of 20 - 95%, non-condensing
- Operation: 20 - 80% rH, storage in ambient relative humidity conditions of 20 - 95%, non-condensing
- Operation: 20 - 80% rH, storage in ambient relative humidity conditions of 20 - 95%, non-condensing
- Operating temperature:
- 10...35 °C
- 10...35 °C
- 10...35 °C
- Protection class:
- IP 31 for optics and electronics
- IP 31 for optics and electronics
- IP 31 for optics and electronics
- Scan:
- Limits freely selectable within the wavelength range; Increment: 0,1/1/5 nm
- Limits freely selectable within the wavelength range; Increment: 0,1/1/5 nm
- Limits freely selectable within the wavelength range; Increment: 0,1/1/5 nm
- Scan speed:
- up to 170 nm/min (depending on the increment)
- up to 750 nm/min (depending on the increment)
- up to 750 nm/min (depending on the increment)
- Spectral bandwidth:
- 4 nm
- 4 nm
- 1,8 nm
- Stray light:
- ≤0,1% transmission at 340 nm
- ≤0,1% transmission at 340 nm; ≤1% transmission at 198 nm
- ≤0,1% transmission at 340 nm; ≤1% transmission at 198 nm
- W×D×H:
- 418×278×169 mm
- 418×278×169 mm
- 418×278×169 mm
- Weight:
- ~6,8 kg
- ~6,8 kg
- ~6,8 kg
- Storage temperature:
- −20...+60 °C for 24 hours
- −20...+60 °C for 24 hours
- −20...+60 °C for 24 hours