VWR® Traceable®, Conductivity meter, handheld
  VWRU23226-505
undefined
VWR® Traceable®, Conductivity meter, handheld
  VWRU23226-505

 

  • Temperature range:
    −30...+130 °C
  • Description:
    Conductivity meter, portable, with probe, Traceable®
  • Environmentally Preferable:
  • W×D×H:
    95×38×171 mm
  • Conductivity accuracy:
    ±0,3% +1 digit
  • Conductivity range:
    0,01 - 20,000 / 0,1 - 200 / 1 - 2000 / 10 - 20000 / 100 - 200000 µS
  • TDS accuracy:
    ±0,3% +1 digit
  • TDS range:
    0,1 - 20,00; 1 - 200; 10 - 2000; 100 - 20000 ppm
  • Weight:
    0,45 kg
  • Pk:
    1

 

 

This portable meter measures the conductivity, resistivity, dissolved solids, salinity and temperature in routine analysis, quality control, and research. Ideal for water analysis, biology, chromatography, food, electronics, dairies, and PC-board rinsing.

  • Auto-ranging feature; automatic and manual temperature compensation
  • Four unique calibration points may be entered into memory utilising solution standards; settings are stored even when unit is off
  • Tough, chemical-resistant ABS housing withstands the harshest environments
  • 12,7 mm (¹/₂") high 4-digit LCD readout shows temperature in °C or °F
  • Recorder jack allows continuous monitoring and a permanent record
  • Conductivity probe has a range of 1,0 to 200000 microhmos

The supplied instant-response probe contains platinum electrodes that deliver highly accurate readings. Internal solid-state thermistor (for automatic/manual temperature compensation) permits all readings to be referenced to the international standard of 25 °C. Temperature compensation is automatic (2% per °C), user-designated (0.000 to 5.000% per °C), or absolute. K-factor may be adjusted to match each probe. Exclusive temperature compensation disable function fulfills USP-NF (United States Pharmacopoeia, National Formulary, 645 Conductivity Measurement) requirement.

Compliance with ISO/IEC 17025:2005 (1750.01); accredited by A2LA indicating traceability to standards provided by NIST.

 : Supplied with glass probe, cable, battery and Traceable® certificate.