Specifications
- Description:Stainless steel sensor
- Length:2.0 mm
Specifications
About this item
Direct quantitative measurements of hydrogen peroxide in biological samples.
Response time: <5 s (90%)
Detection limit: <100 nM to 100 µM
Drift: <0.2 pA/min
Sensitivity: 8 pA/μM
Gas permeable polymer membrane sleeve blocks liquids, ions and particulate matter. Incorporated reference electrode.
Despite the recognized importance of this oxidant in biology, real-time measurements at low concentration have been difficult. They are designed to compliment existing high sensitivity fluorescent approaches with direct quantitative measurement in biological samples in the low nM range.
The ISO HPO-2 is a 2.0 mm stainless steel sensor, with replaceable membrane sleeves and an internal refillable electrolyte. It is designed for use in cell cultures and similar applications.