Gas detection tubes for carbon disulphide
: Dräger
Operating conditions: 0 to 40 °C ; up to 30 mg/l H₂O and for tube 5/a: 3 to 15 mg/l H₂O absolute humidity.
- Data below representative for use with an Accuro® pump
- Cross sensitivity for tube 3/a hydrogen sulphide in the TLV range retained in the pre-layer (does not interfere)
- For tubes carbon disulphide 5/a and 30/a it is impossible to measure carbon disulphide in the presence of hydrogen sulphide
One stroke is equivalent to a volume of 100 ±5 ml of gas passing through the tube.
The standard measuring range is calibrated at 20 °C and 1013 hPa.
- Catalog No:
- 151-0183
- DRAE6728351
- 118-0339
- Model:
- Carbon disiphide 3/a
- Carbon disiphide 5/a
- Carbon disiphide 30/a
- Colour change:
- Pale blue to yellow green
- White to brown green
- Pale blue to brown
- Measurement time:
- 2 minutes
- 3 minutes
- 1 minute
- No. of strokes:
- 15 - 1
- 11
- 6
- Standard deviation:
- ±30%
- ±10 - 15 %
- ±15 - 20 %
- Standard range:
- 3 - 95 ppm
- 5 - 60 ppm
- 0,1 - 10 mg/l