VWR® pHenomenal® pH 1100L/1100LB pH/mV/°C Meter, Bench
This user friendly instrument, with articulated electrode stand and IP 43-rated housing, provides high resolution and accuracy for precise measurements. Complete with USB output for connection to a computer. GLP compliant.
- Large graphic LCD with continuous LED backlight showing both pH/mV and temperature value simultaneously
- Data logging function selectable between 1 min and 1 h; memory capacity of 5000 data sets
- Automatic power shut-off: Adjustable from 10 min to 24 h
- Automatic buffer recognition of DIN and NIST buffers (1,68/4,00/6,86/9,18/12,54)
- Three additional technical buffers at 25 °C (4,00/7,00/10,00) in memory
- Advanced self-test and diagnostics (electrode efficiency diagnostic, drift control); calibration symbol shows when recalibration is required
- Low energy consumption. Choice of mains supply or 4× AA batteries for over 1000 operating hours
- Low battery indicator
: Ready to use pH 1100L kit (662-1657) includes meter glass pH electrode with built-in pHenomenal 221 temperature sensor, 4 LR06 batteries, 1 m cable and articulated stand, 2×50 ml buffers (pH 4,00 and 7,00), 50 ml plastic beaker, short instructions and instruction manual (22 languages) on CD ROM and plugin power supply.
The life science kit pH 1100LB (662-1891) includes the LS 221 Biotech electrode with 1 m cable instead of the 221 electrode.
The life science kit pH 1100LB (662-1891) includes the LS 221 Biotech electrode with 1 m cable instead of the 221 electrode.
- Catalog No:
- 662-1657
- 662-1861
- Model:
- pH 1100L
- pH 1100LB
- pH range:
- −2,000...+20,000
- −2,000...+20,000
- pH resolution:
- 0,1 / 0,01 / 0,001 (selectable)
- 0,1 / 0,01 / 0,001 (selectable)
- pH accuracy:
- ±0,1 / ±0,01 / ±0,005
- ±0,1 / ±0,01 / ±0,005
- pH calibration:
- Up to 5 points
- Up to 5 points
- Automatic buffer recognition:
- VWR® Chemicals TITRINORM® buffers at 25 °C: 4,00 / 6,98 / 9,94
NIST / DIN 19266 (1,68 / 4,00 / 6,86 / 9,18 / 12,54)
Technical buffers at 25 °C: 4,00 / 7,00 / 10,00 - VWR® Chemicals TITRINORM® buffers at 25 °C: 4,00 / 6,98 / 9,94
NIST / DIN 19266 (1,68 / 4,00 / 6,86 / 9,18 / 12,54)
Technical buffers at 25 °C: 4,00 / 7,00 / 10,00
- mV range:
- ±1200,0 / ±2500
- ±1200,0 / ±2500
- mV resolution:
- 0,1 / 1 (selectable)
- 0,1 / 1 (selectable)
- mV accuracy:
- ±0,3 / ±1
- ±0,3 / ±1
- Temperature range:
- Auto: −5,0...+105,0 °C; manual: −25...+130 °C
- Auto: −5,0...+105,0 °C; manual: −25...+130 °C
- Temperature resolution:
- 0,1/1 °C
- 0,1/1 °C
- Temperature accuracy:
- ±0,2 °C
- ±0,2 °C
- Temperature compensation:
- Auto / manual
- Auto / manual
- Display:
- Graphic LCD with LED backlight
- Graphic LCD with LED backlight
- Power supply:
- Mains or 4× AA batteries
- Mains or 4× AA batteries
- Compliance:
- IP 43 / GLP
- IP 43 / GLP
- Data logging - log function:
- Selectable between 1 min and 1 h (selectable intervals: 1 s, 5 s, 10 s, 30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min)
- Selectable between 1 min and 1 h (selectable intervals: 1 s, 5 s, 10 s, 30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min)
- Data logging capacity:
- 5000 data sets
- 5000 data sets
- Connections:
- BNC for pH / banana for temperature
- BNC for pH / banana for temperature
- Outputs:
- USB connection (slave, bidirectional)
- USB connection (slave, bidirectional)
Electrodes