Specifications
- Rango de temperatura:5° above ambient...65 °C (149 °F)
- Intervalo de temperaturas:Ambient +5…65 °C (149 °F)
- Environmentally Preferable:Waste Reducing
- Peso al enviarse:37.6 kg (83 lbs.)
- Tray dimensions:27.9W x 33L cm (11 x 13")
- Max. load:15.9 kg (35 lbs.)
- Overall dimensions:35.6W x 40.6H x 64.8L cm (14 x 16 x 251/2")
- Uniformidad de la temperatura:±0.5°C at 37°C
- Dimensiones de la cámara:30.5W x 24.1H x 34L cm (12 x 91/2 x 1313/32")
- Precisión de la velocidad:±1% of set speed, <100 rpm ±1 rpm
- Temporizador:1 second to 160 hours
- Referencia:12620-946
- Detalles eléctricos:120 V, 50/60 Hz
- Intervalo de velocidad:15 - 500 rpm
- Orbit:19 mm (¾")
Specifications
About this item
The microprocessor control of the Model 3500I provides consistent, uniform mixing action.
- Microprocessor control with digital display for speed, temperature, and time
- Opti-flow forced ventilation system
- Triple eccentric drive with brushless motor
- Five-year warranty
- NIST-Traceable, ISO/IEC 17025 accredited calibration certificate from Troemner available
A spill-resistant touchpad adjusts the temperature in 1 °C increments
For added safety, a ramping feature increases speed to set point, and a built-in load sensor automatically reduces rpm during unbalanced conditions to further prevent spills and protect samples. The brushless motor is designed to handle heavy load, and the spill-resistant design protects internal components from accidental spills and leaks.
The timer displays elapsed time or counts to zero and shuts off the unit when programmed to user-defined limit. An alarm sounds when time reaches zero. The temperature-calibration mode allows the unit to be calibrated to an external temperature device, and the Opti-Flow forced ventilation system uses twin induction fans and air deflectors to deliver temperature uniformity and stability. The Accu-Drive™ Shaking System ensures precise speed control. This system monitors shaking speed and maintains set point even under changing loads.