UV/Vis spectrophotometers, U-3900
634-0754
- Model:U-3900H
- Absorbance range:−5,5 - 5,5 A
- Light source:Tungsten/deuterium
- Description:UV/Vis spectrophotometer
- Environmentally Preferable:
- Scanning speed:1,5; 3; 15; 30; 60; 120; 300; 600; 1200; 1800; 2400 nm/min
- Electrical:100 - 240 V
- W×D×H:680×692×257 mm
- Wavelength range:190 - 900 nm
- Wavelength accuracy:±0,1 nm @ 656 nm after calibration
- Absorbance accuracy:±0,0003 A (190 − 850 nm)
- Cell compartment:120×300×140 mm
- Interfaces:USB
- Optical design:Double beam (100 mm spacing) with double monochromator
- Photometric accuracy:±0,3% T
- Photometric range:0 - 300% T
- Spectral bandwidth:0,1; 0,5; 1; 2; 4; 5 nm − 6 steps
- Stray light:0,00025% (NaI: 220 nm, NaNO₂: 340 nm)
- Weight:45 kg
- Pk:1
U-3900 and U-3900H spectrophotometers are comprehensive, research-grade solutions. They provide excellent performance afforded by low stray light, selectable band pass and high energy.
- Double-beam stability provides improved accuracy and repeatability over extended periods of time
- Ability to reference-beam attenuate highly absorbing species for improved signal to noise
- Selectable scanning rate range of 1,5 to 2400 nm/min and changeable scan rate for the UV region to minimise noise further
- Interface via USB and a laptop or notebook PC format, can easily sit on a flat surface of the instrument (see picture)
The U-3900H is useful for measuring of samples with high absorbance/low transmittance yet still achieving excellent linearity.
: Including sipper systems, alternative cell holders, filter holders and integrating spheres. Please contact your local Avantor specialists.
: UV Solutions for U-3900 software (supplied) offers the flexibility and wide ranging functionality.