Drying and Heating Chambers, with Forced Convection, M Avantgarde.Line
Supplier: Binder
Certificates
About this item
The M Series Avantgarde.Line units have a maximum of 300 °C and various advanced programming options ensure these material test chambers are ideal for material and aging tests. The units have a high air flow and their powerful fans ensure rapid heating.
- Temperature range: 15 to 22 °C above ambient temperature up to +300 °C
- High temperature accuracy thanks to APT.line™ technology with increased heating output
- Reinforced fan with adjustable speed
- Electromechanical control of the exhaust air flap
- Intuitive touchscreen controller with time-segment and real-time programming
- Class 2 independent adjustable temperature safety device (DIN 12880) and class 3.1 independent temperature safety device (DIN 12880) with visual/acoustic alarm available for selection on the controller
- Failsafe function - backup sensor PT100
- Computer interface: Ethernet
Units have an internal data logger, the measured values can be read out in open format via USB.
Heat-up time: 10 min to 150 °C (M 56), 13 min to 150 °C (M 115), 14 min to 150 °C (M 260 and M 720).
Certifications: IP protection class according to DIN EN 60529: IP 20.
Delivery: Supplied with two chrome plated racks.
Specifications
- Catalog No:
- 466-0703
- 466-0704
- 466-0702
- 466-0701
- Model:
- M 260
- M 720
- M 115
- M 56
- Temperature range:
- Ambient 22…300 °C
- Ambient 22…300 °C
- Ambient 22…300 °C
- Ambient 22…300 °C
- Temperature fluctuation (time):
- ±0,3 K @ 150 °C
- ±0,5 K @ 150 °C
- ±0,2 K @ 150 °C
- ±0,2 K @ 150 °C
- Temperature variation (spatial):
- ±1,7 K @ 150 °C
- ±2 K @ 150 °C
- ±1,4 K @ 150 °C
- ±1,4 K @ 150 °C
- Capacity:
- 256 L
- 735 L
- 113 L
- 59 L
- Convection type:
- Forced convection
- Forced convection
- Forced convection
- Forced convection
- Int. W×D×H:
- 650×505×780 mm
- 1000×565×1300 mm
- 550×375×550 mm
- 400×335×440 mm
- Ext. W×D×H:
- 810×786×962 mm
- 1160×816×1590 mm
- 710×636×732 mm
- 560×596×622 mm
- Max. load per shelf:
- 40 kg/shelf; 270 kg/unit
- 45 kg/shelf; 315 kg/unit
- 30 kg/shelf; 150 kg/unit
- 30 kg/shelf; 70 kg/unit
- Power consumption:
- 3,1 kW
- 6 kW
- 2,4 kW
- 2 kW
- Recovery Time:
- 7 min after 30 s opening at 150 °C
- 7 min after 30 s opening at 150 °C
- 7 min after 30 s opening at 150 °C
- 4 min after 30 s opening at 150 °C
- Shelves supplied (max.):
- 2 (8)
- 2 (16)
- 2 (4)
- 2 (4)
- Weight:
- 85 kg
- 169 kg
- 56 kg
- 43 kg
Specifications
Options
Product Information
Frequently Bought Together