Multi-parameter meter (pH/ISE/conductivity/oxygen), bench top, Orion™ Versa Star Pro™
These bench top meters with pH/ISE, conductivity and Dissolved Oxygen/RDO modules meet the most challenging applications for pH, ion concentration, conductivity, dissolved oxygen and temperature. The meters offer interchangeable measurement modules that allow multiple users to customise four separate channels to meet their specific requirements.
- Colour display with adjustable text sizes; numeric keypad; onscreen measurement stability indicator
- Auto-Read to hold a stable reading; ten password-protected methods per channel
- Audible alarms automatically notify users when a set offset value, high/low limit value, and calibration time interval are reached
- Uniformly mix solutions using up to two meter-controlled stirrer probes that are easy to position and can be quickly rinsed between samples
- Data log collects and date/time stamps up to 2000 measurement sets with optional electrode ID, sample ID and account ID
- Calibration log stores ten most recent calibrations per parameter with viewable graph for multipoint pH and ISE calibrations
- Meter facilitates data transfers to a computer via USB or RS232 connections
IP 54-rated housing resists dust and splashes and can be wall-mounted if table space is limited.
When used in conjunction with the Orion Navigator Pro software, the Orion Versa Star Pro meter is designed to create electronic data records and audit trails with password-protection, which can be used to assist with 21 CFR Part 11 requirements when paired with each laboratory’s proper IT protocols, SOPs and training records.*Orion Navigator Pro software offers multi-level system access for an administrator and up to 10 users. The system administrator can customize each user’s access level from full system access to basic data collection only. The Navigator Pro software controls all meter functions including meter settings, calibration, measurement, and reporting to help ensure data integrity and provide a complete audit trail.
Features and benefits
- Generates electronic records and signatures
- Creates audit trail report for electronic records
- Limits access to authorized individuals
- Password protection with system administrator and up to 10 users
- Full computer control of meter functions including setup, calibration and measurement
- Paperless record keeping system
: All meters are supplied with one or two VSTAR-ISE pH/ISE modules, one VSTAR-CND conductivity module, one VSTAR-RD RDO/DO module, electrode stand, universal power adapter, USB cable, literature CD, printed quick start guide, computer interface cable and meter test certificate.
Kits additionally include: 8157BNUMD ROSS Ultra Triode epoxy-body pH/ATC electrode, 013005MD Orion DuraProbe 4-cell (K=0,475) conductivity sensor, 083005MD Orion polarographic dissolved oxygen sensor, 096019 Orion Star stirrer probe, 810199 ROSS pH buffer and solution kit, 011007 Orion 1413 µS conductivity standard, 080513 DO maintenance kit, BOD adapter/funnel/stirrer.
: * It is the responsibility of each laboratory to validate the Navigator Pro computer software for its intended use, including all regulatory requirements.
These bench top meters with pH/ISE, conductivity and Dissolved Oxygen/RDO modules meet the most challenging applications for pH, ion concentration, conductivity, dissolved oxygen and temperature. The meters offer interchangeable measurement modules that allow multiple users to customise four separate channels to meet their specific requirements.
- Colour display with adjustable text sizes; numeric keypad; onscreen measurement stability indicator
- Auto-Read to hold a stable reading; ten password-protected methods per channel
- Audible alarms automatically notify users when a set offset value, high/low limit value, and calibration time interval are reached
- Uniformly mix solutions using up to two meter-controlled stirrer probes that are easy to position and can be quickly rinsed between samples
- Data log collects and date/time stamps up to 2000 measurement sets with optional electrode ID, sample ID and account ID
- Calibration log stores ten most recent calibrations per parameter with viewable graph for multipoint pH and ISE calibrations
- Meter facilitates data transfers to a computer via USB or RS232 connections
IP 54-rated housing resists dust and splashes and can be wall-mounted if table space is limited.
When used in conjunction with the Orion Navigator Pro software, the Orion Versa Star Pro meter is designed to create electronic data records and audit trails with password-protection, which can be used to assist with 21 CFR Part 11 requirements when paired with each laboratory’s proper IT protocols, SOPs and training records.*Orion Navigator Pro software offers multi-level system access for an administrator and up to 10 users. The system administrator can customize each user’s access level from full system access to basic data collection only. The Navigator Pro software controls all meter functions including meter settings, calibration, measurement, and reporting to help ensure data integrity and provide a complete audit trail.
Features and benefits
- Generates electronic records and signatures
- Creates audit trail report for electronic records
- Limits access to authorized individuals
- Password protection with system administrator and up to 10 users
- Full computer control of meter functions including setup, calibration and measurement
- Paperless record keeping system
: All meters are supplied with one or two VSTAR-ISE pH/ISE modules, one VSTAR-CND conductivity module, one VSTAR-RD RDO/DO module, electrode stand, universal power adapter, USB cable, literature CD, printed quick start guide, computer interface cable and meter test certificate.
Kits additionally include: 8157BNUMD ROSS Ultra Triode epoxy-body pH/ATC electrode, 013005MD Orion DuraProbe 4-cell (K=0,475) conductivity sensor, 083005MD Orion polarographic dissolved oxygen sensor, 096019 Orion Star stirrer probe, 810199 ROSS pH buffer and solution kit, 011007 Orion 1413 µS conductivity standard, 080513 DO maintenance kit, BOD adapter/funnel/stirrer.
: * It is the responsibility of each laboratory to validate the Navigator Pro computer software for its intended use, including all regulatory requirements.
- Catalog No:
- ORIOVSTAR90
- ORIOVSTAR91
- ORIOVSTAR92
- ORIOVSTAR93
- Model:
- Orion™ Versa Star Pro™
- Orion™ Versa Star Pro™
- Orion™ Versa Star Pro™
- VERSA STAR™ 91/93
- pH - Område:
- -2...+20
- -2...+20
- -2...+20
- -2...+20
- pH - Aflæsningsinterval:
- 0,1/0,01/0,001
- 0,1/0,01/0,001
- 0,1/0,01/0,001
- 0,1/0,01/0,001
- pH - Nøjagtighed:
- ±0,002
- ±0,002
- ±0,002
- ±0,002
- pH - Kalibrering:
- op til 6 punkter
- op til 6 punkter
- op til 6 punkter
- op til 6 punkter
- mV - Område:
- ±2.000
- ±2.000
- ±2.000
- ±2.000
- mV - Aflæsningsinterval:
- 0,1
- 0,1
- 0,1
- 0,1
- mV - Nøjagtighed:
- ±0,2 mV eller ±0,05 % af udlæsningen, alt efter hvad der er størst.
- ±0,2 mV eller ±0,05 % af udlæsningen, alt efter hvad der er størst.
- ±0,2 mV eller ±0,05 % af udlæsningen, alt efter hvad der er størst.
- ±0,2 mV eller ±0,05 % af udlæsningen, alt efter hvad der er størst.
- mV - Kalibrering:
- 1 point
- 1 point
- 1 point
- 1 point
- ISE Område:
- 0 - 19999
- 0,0001 - 19990 ppm
- 0,0001 - 19990 ppm
- 0 - 19999
- ISE Opløsning:
- 0,0001 minimum, 1 to 4 significant digits (user selectable)
- 0,0001 minimum, 1 to 4 significant digits (user selectable)
- 0,0001 minimum, 1 to 4 significant digits (user selectable)
- op til 4 signifikante cifre
- ISE Nøjagtighed:
- ±0,2 mV eller ±0,5 %, alt efter hvad der er størst
- ±0,2 mV or ±0,05% of reading (whichever is greater)
- ±0,2 mV eller ±0,5 %, alt efter hvad der er størst
- ±0,2 mV eller ±0,5 %, alt efter hvad der er størst
- ISE Kalibrering:
- op til 6
- op til 6
- op til 6
- op til 6
- ISE Enheder:
- ppm, M, mg/L, %, ppb eller uden enhed
- ppm, M, mg/L, %, ppb eller uden enhed
- ppm, M, mg/L, %, ppb eller uden enhed
- ppm, M, mg/L, %, ppb eller uden enhed
- Konduktivitetsområde:
- 0,00 - 3.000 mS/cm
- 0,00 - 3.000 mS/cm
- 0,001 µS/cm - 3000 mS/cm
- 0,00 - 3000 mS/cm
- Konduktivitetsopløsning:
- 0,01 µS minimum; minimum 4 signifikante tal
- 0,01 µS minimum; minimum 4 signifikante tal
- 0,01 µS minimum; minimum 4 signifikante tal
- 0,01 µS minimum; minimum 4 signifikante tal
- Konduktivitetsnøjagtighed:
- ±0,5 % af udlæsningen ±1 ciffer
- 0,5% of reading ±1 digit (> 3 μS); 0,5% of reading ±0,01 μS (≤ 3 μS)
- 0,5% of reading ±1 digit (> 3 μS); 0,5% of reading ±0,01 μS (≤ 3 μS)
- 0,5% of reading ±1 digit (> 3 μS); 0,5% of reading ±0,01 μS (≤ 3 μS)
- Konduktivitetscellekonstant:
- 0,001 - 199,9 cm⁻¹
- 0,001 - 199,9 cm⁻¹
- 0,001 - 199,9 cm⁻¹
- 0,001 - 199,9 cm⁻¹
- Konduktivitetskalibrering:
- op til 6
- op til 6
- op til 6
- op til 6
- Modstand - Område:
- 2 Ω/cm - 100,0 MΩ/cm
- 2 Ω/cm - 100,0 MΩ/cm
- 2 Ω/cm - 100,0 MΩ/cm
- 2 Ω/cm - 100,0 MΩ/cm
- Modstand - Nøjagtighed:
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- Resistivity resolution:
- 1 Ω/cm/0,1 MΩ/cm
- 1 Ω/cm/0,1 MΩ/cm
- 1 Ω/cm/0,1 MΩ/cm
- 1 Ω/cm/0,1 MΩ/cm
- Salinitet - Område:
- 0,06 - 80,00 psu/0,05 - 42,00 ppt
- 0,06 - 80,00 psu/0,05 - 42,00 ppt
- 0,06 - 80,00 psu/0,05 - 42,00 ppt
- 0,06 - 80,00 psu/0,05 - 42,00 ppt
- Salinitet - Aflæsningsinterval:
- 0,01 psu or 0,01 ppt minimum, auto ranging
- 0,01 psu or 0,01 ppt minimum, auto ranging
- 0,01 psu or 0,01 ppt minimum, auto ranging
- 0,01 psu or 0,01 ppt minimum, auto ranging
- Salinitet - Nøjagtighed:
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- TDS - Område:
- 0,001 - 200,0 ppm
- 0,001 - 200,0 ppm
- 0,001 - 200,0 ppm
- 0,001 - 200,0 ppm
- TDS - Aflæsningsinterval:
- 0,001 ppm minimum, auto ranging up to 4 significant digits
- 0,001 ppm minimum, auto ranging up to 4 significant digits
- 0,001 ppm minimum, auto ranging up to 4 significant digits
- 0,001 ppm minimum, auto ranging up to 4 significant digits
- TDS - Nøjagtighed:
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- TDS - Omregningsfaktor:
- Linear (0,02 - 9,99)
- Linear (0,02 - 9,99)
- Linear (0,02 - 9,99)
- Linear (0,02 - 9,99)
- DO Koncentration:
- DO: 0,00 - 90,00 mg/L; RDO: 0,00 - 50,00 mg/L
- Polarografisk: 0 - 90 mg/l
RDO optisk: 0 - 50 mg/l - DO: 0,00 - 90,00 mg/L; RDO: 0,00 - 50,00 mg/L
- DO: 0,00 - 90,00 mg/L; RDO: 0,00 - 50,00 mg/L
- D.O. - Aflæsningsinterval:
- 0,01/0,1 mg/L
- 0,01/0,1 mg/L
- 0,01/0,1 mg/L
- 0,01/0,1 mg/L
- D.O. - mætning:
- 0,0 - 600,0 %
- 0,0 - 600,0 %
- 0,0 - 600,0 %
- 0,0 - 600,0 %
- D.O. Nøjagtighed:
- DO: ±0,2 mg/l or ±2% of reading (whichever is greater)
RDO: ±0,1 mg/l (up to 8 mg/l), ±0,2 mg/l (8 - 20 mg/l), ±10% of reading (up to 50 mg/l)
±2 % saturation or ±2% of reading whichever is greater - Polarografisk: ±0,2 mg/l; ±2 %
RDO optisk: ±0,1 mg/l (op til 8 mg/l), ±0,2 mg/l (op til 20 mg/l), 10 % af udlæsningen herudover; ±2 % - DO: ±0,2 mg/l or ±2% of reading (whichever is greater)
RDO: ±0,1 mg/l (up to 8 mg/l), ±0,2 mg/l (8 - 20 mg/l), ±10% of reading (up to 50 mg/l)
±2 % saturation or ±2% of reading whichever is greater - DO: ±0,2 mg/l or ±2% of reading (whichever is greater)
RDO: ±0,1 mg/l (up to 8 mg/l), ±0,2 mg/l (8 - 20 mg/l), ±10% of reading (up to 50 mg/l)
±2 % saturation or ±2% of reading whichever is greater
- D.O. - Barometertrykområde:
- 400 - 850 mmHg
- 400 - 850 mmHg
- 400 - 850 mmHg
- 400 - 850 mmHg
- Temperaturinterval:
- −5,0...+105,0 °C
- −5,0...+105,0 °C
- −5,0...+105,0 °C
- −5,0...+105,0 °C
- Temperature resolution:
- 0,1 °C
- 0,1 °C
- 0,1 °C
- 0,1 °C
- Temperature accuracy:
- ±0,1 °C
- ±0,1 °C
- ±0,1 °C
- ±0,1 °C
- Temperaturkalibrering:
- 1 point offset
- 1 point offset
- 1 point offset
- 1 point offset
- Temperaturkompensation:
- Linear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or off
- Linear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or off
- Linear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or off
- Linear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or off
- Display:
- Farve, grafisk LCD
- Farve, grafisk LCD
- Colour graphic LCD with backlight
- Colour graphic LCD with backlight
- Strømforsyning:
- Universal AC adapter, 100 - 240 VAC, 0 - 60 Hz,
- Universal, AC-strømadapter, 90 - 260 VAC, 50 - 60 Hz
- Universal, AC-strømadapter, 90 - 260 VAC, 50 - 60 Hz
- Universal, AC-strømadapter, 90 - 260 VAC, 50 - 60 Hz
- Overensstemmelse:
- CE, TUV, FCC Klasse A, grænser
- IP 54
- IP 54
- IP 54
- Operating temperature range:
- 5...45 °C (ambient)
- 5...45 °C (ambient)
- 5...45 °C (ambient)
- 5...45 °C (ambient)
- Datalog - Log-funktion:
- Automatic data logging with Auto-Read, Timed and Single Shot measure modes; manual data logging with Continuous measure mode
- Automatic data logging with Auto-Read, Timed and Single Shot measure modes; manual data logging with Continuous measure mode
- Automatic data logging with Auto-Read, Timed and Single Shot measure modes; manual data logging with Continuous measure mode
- Automatic data logging with Auto-Read, Timed and Single Shot measure modes; manual data logging with Continuous measure mode
- Datalog Kapacitet:
- 2.000 punkter
- 2.000 punkter
- 2.000 punkter
- 2.000 punkter
- Sprog:
- English, Spanish, French, Italian, German, Chinese, Portuguese and Korean
- English, Spanish, French, Italian, German, Chinese, Portuguese and Korean
- English, Spanish, French, Italian, German, Chinese, Portuguese and Korean
- English, Spanish, French, Italian, German, Chinese, Portuguese and Korean
- Tilslutninger:
- BNC (pH or ISE), 8-pin Mini-DIN (temperature), reference pin-tip, 8-pin Mini-DIN (conductivity with temperature probe), 9-pin Mini-DIN (RDO or polarographic DO with temperature probe), 2 stirrer inputs
- BNC (pH or ISE), 8-pin Mini-DIN (temperature), reference pin-tip, 8-pin Mini-DIN (conductivity with temperature probe), 9-pin Mini-DIN (RDO or polarographic DO with temperature probe), 2 stirrer inputs
- BNC (pH or ISE), 8-pin Mini-DIN (temperature), reference pin-tip, 8-pin Mini-DIN (conductivity with temperature probe), 9-pin Mini-DIN (RDO or polarographic DO with temperature probe), 2 stirrer inputs
- BNC (pH or ISE), 8-pin Mini-DIN (temperature), reference pin-tip, 8-pin Mini-DIN (conductivity with temperature probe), 9-pin Mini-DIN (RDO or polarographic DO with temperature probe), 2 stirrer inputs
- Output:
- RS232, USB (2)
- RS-232, USB
- RS-232, USB
- RS-232, USB
- Antal kanaler:
- 4 - customisable with modules
- 4 - customisable with modules
- 4 - customisable with modules
- 4 - customisable with modules
- Kalibrering:
- pH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ
- pH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ
- pH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ
- pH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ
- Kalibreringspunkter:
- 1 point relative mV (RmV) mode, 1 point ORP mode
- 1 point relative mV (RmV) mode, 1 point ORP mode
- 1 point relative mV (RmV) mode, 1 point ORP mode
- 1 point relative mV (RmV) mode, 1 point ORP mode
- B×D×H:
- 280×240×100 mm
- 280×240×100 mm
- 280×240×100 mm
- 280×240×100 mm
- Vægt:
- 1,2 kg (modules add 0,2 kg each)
- 1,2 kg (modules add 0,2 kg each)
- 1,2 kg (modules add 0,2 kg each)
- 1,2 kg (modules add 0,2 kg each)
Product Family Options
- Beskrivelse
- Multi-parameter meter, bench top, Orion™ Versa Star™ Pro™ meter, with 1 × pH/ISE, 1 × conductivity and 1 × RDO/DO module (without electrodes)1 * 1 Items40.590,00 kr
Specifications
Cat. No.ORIOVSTAR90ModelOrion™ Versa Star Pro™pH - Område-2...+20pH - Aflæsningsinterval0,1/0,01/0,001pH - Nøjagtighed±0,002pH - Kalibreringop til 6 punktermV - Område±2.000mV - Aflæsningsinterval0,1mV - Nøjagtighed±0,2 mV eller ±0,05 % af udlæsningen, alt efter hvad der er størst.mV - Kalibrering1 pointISE Område0 - 19999ISE Opløsning0,0001 minimum, 1 to 4 significant digits (user selectable)ISE Nøjagtighed±0,2 mV eller ±0,5 %, alt efter hvad der er størstISE Kalibreringop til 6ISE Enhederppm, M, mg/L, %, ppb eller uden enhedKonduktivitetsområde0,00 - 3.000 mS/cmKonduktivitetsopløsning0,01 µS minimum; minimum 4 signifikante talKonduktivitetsnøjagtighed±0,5 % af udlæsningen ±1 cifferKonduktivitetscellekonstant0,001 - 199,9 cm⁻¹Konduktivitetskalibreringop til 6Modstand - Område2 Ω/cm - 100,0 MΩ/cmModstand - Nøjagtighed0,5% of reading ±1 digitResistivity resolution1 Ω/cm/0,1 MΩ/cmSalinitet - Område0,06 - 80,00 psu/0,05 - 42,00 pptSalinitet - Aflæsningsinterval0,01 psu or 0,01 ppt minimum, auto rangingSalinitet - Nøjagtighed0,5% of reading ±1 digitTDS - Område0,001 - 200,0 ppmTDS - Aflæsningsinterval0,001 ppm minimum, auto ranging up to 4 significant digitsTDS - Nøjagtighed0,5% of reading ±1 digitTDS - OmregningsfaktorLinear (0,02 - 9,99)DO KoncentrationDO: 0,00 - 90,00 mg/L; RDO: 0,00 - 50,00 mg/LD.O. - Aflæsningsinterval0,01/0,1 mg/LD.O. - mætning0,0 - 600,0 %D.O. NøjagtighedDO: ±0,2 mg/l or ±2% of reading (whichever is greater)
RDO: ±0,1 mg/l (up to 8 mg/l), ±0,2 mg/l (8 - 20 mg/l), ±10% of reading (up to 50 mg/l)
±2 % saturation or ±2% of reading whichever is greaterD.O. - Barometertrykområde400 - 850 mmHgTemperaturinterval−5,0...+105,0 °CTemperature resolution0,1 °CTemperature accuracy±0,1 °CTemperaturkalibrering1 point offsetTemperaturkompensationLinear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or offDisplayFarve, grafisk LCDStrømforsyningUniversal AC adapter, 100 - 240 VAC, 0 - 60 Hz,OverensstemmelseCE, TUV, FCC Klasse A, grænserOperating temperature range5...45 °C (ambient)Datalog - Log-funktionAutomatic data logging with Auto-Read, Timed and Single Shot measure modes; manual data logging with Continuous measure modeDatalog Kapacitet2.000 punkterSprogEnglish, Spanish, French, Italian, German, Chinese, Portuguese and KoreanTilslutningerBNC (pH or ISE), 8-pin Mini-DIN (temperature), reference pin-tip, 8-pin Mini-DIN (conductivity with temperature probe), 9-pin Mini-DIN (RDO or polarographic DO with temperature probe), 2 stirrer inputsOutputRS232, USB (2)Antal kanaler4 - customisable with modulesKalibreringpH with calibration editing option, ion concentration (ISE) with calibration editing option, relativKalibreringspunkter1 point relative mV (RmV) mode, 1 point ORP modeB×D×H280×240×100 mmVægt1,2 kg (modules add 0,2 kg each) - Multi-parameter meter, bench top, Orion™ Versa Star™ Pro™ meter, with 2 × pH/ISE, 1 × conductivity and 1 × RDO/DO module (without electrodes)1 * 1 Items50.420,00 kr
Specifications
Cat. No.ORIOVSTAR91ModelOrion™ Versa Star Pro™pH - Område-2...+20pH - Aflæsningsinterval0,1/0,01/0,001pH - Nøjagtighed±0,002pH - Kalibreringop til 6 punktermV - Område±2.000mV - Aflæsningsinterval0,1mV - Nøjagtighed±0,2 mV eller ±0,05 % af udlæsningen, alt efter hvad der er størst.mV - Kalibrering1 pointISE Område0,0001 - 19990 ppmISE Opløsning0,0001 minimum, 1 to 4 significant digits (user selectable)ISE Nøjagtighed±0,2 mV or ±0,05% of reading (whichever is greater)ISE Kalibreringop til 6ISE Enhederppm, M, mg/L, %, ppb eller uden enhedKonduktivitetsområde0,00 - 3.000 mS/cmKonduktivitetsopløsning0,01 µS minimum; minimum 4 signifikante talKonduktivitetsnøjagtighed0,5% of reading ±1 digit (> 3 μS); 0,5% of reading ±0,01 μS (≤ 3 μS)Konduktivitetscellekonstant0,001 - 199,9 cm⁻¹Konduktivitetskalibreringop til 6Modstand - Område2 Ω/cm - 100,0 MΩ/cmModstand - Nøjagtighed0,5% of reading ±1 digitResistivity resolution1 Ω/cm/0,1 MΩ/cmSalinitet - Område0,06 - 80,00 psu/0,05 - 42,00 pptSalinitet - Aflæsningsinterval0,01 psu or 0,01 ppt minimum, auto rangingSalinitet - Nøjagtighed0,5% of reading ±1 digitTDS - Område0,001 - 200,0 ppmTDS - Aflæsningsinterval0,001 ppm minimum, auto ranging up to 4 significant digitsTDS - Nøjagtighed0,5% of reading ±1 digitTDS - OmregningsfaktorLinear (0,02 - 9,99)DO KoncentrationPolarografisk: 0 - 90 mg/l
RDO optisk: 0 - 50 mg/lD.O. - Aflæsningsinterval0,01/0,1 mg/LD.O. - mætning0,0 - 600,0 %D.O. NøjagtighedPolarografisk: ±0,2 mg/l; ±2 %
RDO optisk: ±0,1 mg/l (op til 8 mg/l), ±0,2 mg/l (op til 20 mg/l), 10 % af udlæsningen herudover; ±2 %D.O. - Barometertrykområde400 - 850 mmHgTemperaturinterval−5,0...+105,0 °CTemperature resolution0,1 °CTemperature accuracy±0,1 °CTemperaturkalibrering1 point offsetTemperaturkompensationLinear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or offDisplayFarve, grafisk LCDStrømforsyningUniversal, AC-strømadapter, 90 - 260 VAC, 50 - 60 HzOverensstemmelseIP 54Operating temperature range5...45 °C (ambient)Datalog - Log-funktionAutomatic data logging with Auto-Read, Timed and Single Shot measure modes; manual data logging with Continuous measure modeDatalog Kapacitet2.000 punkterSprogEnglish, Spanish, French, Italian, German, Chinese, Portuguese and KoreanTilslutningerBNC (pH or ISE), 8-pin Mini-DIN (temperature), reference pin-tip, 8-pin Mini-DIN (conductivity with temperature probe), 9-pin Mini-DIN (RDO or polarographic DO with temperature probe), 2 stirrer inputsOutputRS-232, USBAntal kanaler4 - customisable with modulesKalibreringpH with calibration editing option, ion concentration (ISE) with calibration editing option, relativKalibreringspunkter1 point relative mV (RmV) mode, 1 point ORP modeB×D×H280×240×100 mmVægt1,2 kg (modules add 0,2 kg each) - Multi-parameter meter, bench top, Orion™ Versa Star™ Pro™ kit, with 1 × pH/ISE, 1 × conductivity and 1 × RDO/DO module1 * 1 Items51.230,00 kr
Specifications
Cat. No.ORIOVSTAR92ModelOrion™ Versa Star Pro™pH - Område-2...+20pH - Aflæsningsinterval0,1/0,01/0,001pH - Nøjagtighed±0,002pH - Kalibreringop til 6 punktermV - Område±2.000mV - Aflæsningsinterval0,1mV - Nøjagtighed±0,2 mV eller ±0,05 % af udlæsningen, alt efter hvad der er størst.mV - Kalibrering1 pointISE Område0,0001 - 19990 ppmISE Opløsning0,0001 minimum, 1 to 4 significant digits (user selectable)ISE Nøjagtighed±0,2 mV eller ±0,5 %, alt efter hvad der er størstISE Kalibreringop til 6ISE Enhederppm, M, mg/L, %, ppb eller uden enhedKonduktivitetsområde0,001 µS/cm - 3000 mS/cmKonduktivitetsopløsning0,01 µS minimum; minimum 4 signifikante talKonduktivitetsnøjagtighed0,5% of reading ±1 digit (> 3 μS); 0,5% of reading ±0,01 μS (≤ 3 μS)Konduktivitetscellekonstant0,001 - 199,9 cm⁻¹Konduktivitetskalibreringop til 6Modstand - Område2 Ω/cm - 100,0 MΩ/cmModstand - Nøjagtighed0,5% of reading ±1 digitResistivity resolution1 Ω/cm/0,1 MΩ/cmSalinitet - Område0,06 - 80,00 psu/0,05 - 42,00 pptSalinitet - Aflæsningsinterval0,01 psu or 0,01 ppt minimum, auto rangingSalinitet - Nøjagtighed0,5% of reading ±1 digitTDS - Område0,001 - 200,0 ppmTDS - Aflæsningsinterval0,001 ppm minimum, auto ranging up to 4 significant digitsTDS - Nøjagtighed0,5% of reading ±1 digitTDS - OmregningsfaktorLinear (0,02 - 9,99)DO KoncentrationDO: 0,00 - 90,00 mg/L; RDO: 0,00 - 50,00 mg/LD.O. - Aflæsningsinterval0,01/0,1 mg/LD.O. - mætning0,0 - 600,0 %D.O. NøjagtighedDO: ±0,2 mg/l or ±2% of reading (whichever is greater)
RDO: ±0,1 mg/l (up to 8 mg/l), ±0,2 mg/l (8 - 20 mg/l), ±10% of reading (up to 50 mg/l)
±2 % saturation or ±2% of reading whichever is greaterD.O. - Barometertrykområde400 - 850 mmHgTemperaturinterval−5,0...+105,0 °CTemperature resolution0,1 °CTemperature accuracy±0,1 °CTemperaturkalibrering1 point offsetTemperaturkompensationLinear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or offDisplayColour graphic LCD with backlightStrømforsyningUniversal, AC-strømadapter, 90 - 260 VAC, 50 - 60 HzOverensstemmelseIP 54Operating temperature range5...45 °C (ambient)Datalog - Log-funktionAutomatic data logging with Auto-Read, Timed and Single Shot measure modes; manual data logging with Continuous measure modeDatalog Kapacitet2.000 punkterSprogEnglish, Spanish, French, Italian, German, Chinese, Portuguese and KoreanTilslutningerBNC (pH or ISE), 8-pin Mini-DIN (temperature), reference pin-tip, 8-pin Mini-DIN (conductivity with temperature probe), 9-pin Mini-DIN (RDO or polarographic DO with temperature probe), 2 stirrer inputsOutputRS-232, USBAntal kanaler4 - customisable with modulesKalibreringpH with calibration editing option, ion concentration (ISE) with calibration editing option, relativKalibreringspunkter1 point relative mV (RmV) mode, 1 point ORP modeB×D×H280×240×100 mmVægt1,2 kg (modules add 0,2 kg each) - Multi-parameter meter, bench top, Orion™ Versa Star™ Pro™ kit, with 2 × pH/ISE, 1 × conductivity and 1 × RDO/DO module1 * 1 Items61.870,00 kr
Specifications
Cat. No.ORIOVSTAR93ModelVERSA STAR™ 91/93pH - Område-2...+20pH - Aflæsningsinterval0,1/0,01/0,001pH - Nøjagtighed±0,002pH - Kalibreringop til 6 punktermV - Område±2.000mV - Aflæsningsinterval0,1mV - Nøjagtighed±0,2 mV eller ±0,05 % af udlæsningen, alt efter hvad der er størst.mV - Kalibrering1 pointISE Område0 - 19999ISE Opløsningop til 4 signifikante cifreISE Nøjagtighed±0,2 mV eller ±0,5 %, alt efter hvad der er størstISE Kalibreringop til 6ISE Enhederppm, M, mg/L, %, ppb eller uden enhedKonduktivitetsområde0,00 - 3000 mS/cmKonduktivitetsopløsning0,01 µS minimum; minimum 4 signifikante talKonduktivitetsnøjagtighed0,5% of reading ±1 digit (> 3 μS); 0,5% of reading ±0,01 μS (≤ 3 μS)Konduktivitetscellekonstant0,001 - 199,9 cm⁻¹Konduktivitetskalibreringop til 6Modstand - Område2 Ω/cm - 100,0 MΩ/cmModstand - Nøjagtighed0,5% of reading ±1 digitResistivity resolution1 Ω/cm/0,1 MΩ/cmSalinitet - Område0,06 - 80,00 psu/0,05 - 42,00 pptSalinitet - Aflæsningsinterval0,01 psu or 0,01 ppt minimum, auto rangingSalinitet - Nøjagtighed0,5% of reading ±1 digitTDS - Område0,001 - 200,0 ppmTDS - Aflæsningsinterval0,001 ppm minimum, auto ranging up to 4 significant digitsTDS - Nøjagtighed0,5% of reading ±1 digitTDS - OmregningsfaktorLinear (0,02 - 9,99)DO KoncentrationDO: 0,00 - 90,00 mg/L; RDO: 0,00 - 50,00 mg/LD.O. - Aflæsningsinterval0,01/0,1 mg/LD.O. - mætning0,0 - 600,0 %D.O. NøjagtighedDO: ±0,2 mg/l or ±2% of reading (whichever is greater)
RDO: ±0,1 mg/l (up to 8 mg/l), ±0,2 mg/l (8 - 20 mg/l), ±10% of reading (up to 50 mg/l)
±2 % saturation or ±2% of reading whichever is greaterD.O. - Barometertrykområde400 - 850 mmHgTemperaturinterval−5,0...+105,0 °CTemperature resolution0,1 °CTemperature accuracy±0,1 °CTemperaturkalibrering1 point offsetTemperaturkompensationLinear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or offDisplayColour graphic LCD with backlightStrømforsyningUniversal, AC-strømadapter, 90 - 260 VAC, 50 - 60 HzOverensstemmelseIP 54Operating temperature range5...45 °C (ambient)Datalog - Log-funktionAutomatic data logging with Auto-Read, Timed and Single Shot measure modes; manual data logging with Continuous measure modeDatalog Kapacitet2.000 punkterSprogEnglish, Spanish, French, Italian, German, Chinese, Portuguese and KoreanTilslutningerBNC (pH or ISE), 8-pin Mini-DIN (temperature), reference pin-tip, 8-pin Mini-DIN (conductivity with temperature probe), 9-pin Mini-DIN (RDO or polarographic DO with temperature probe), 2 stirrer inputsOutputRS-232, USBAntal kanaler4 - customisable with modulesKalibreringpH with calibration editing option, ion concentration (ISE) with calibration editing option, relativKalibreringspunkter1 point relative mV (RmV) mode, 1 point ORP modeB×D×H280×240×100 mmVægt1,2 kg (modules add 0,2 kg each)
Modules
- Beskrivelse