Ordreindtastning
Denmark
Orders LinkContactUsLinkComponent
 

Microplate reader, single-mode, SpectraMax® ABS

Microplate reader, single-mode, SpectraMax® ABS
 
Microplate reader, single-mode, SpectraMax® ABS
Microplate reader, single-mode, SpectraMax® ABS

 

 

SpectraMax® ABS microplate reader bridges the gap between the affordability of filter-based readers and the flexibility of monochromator-based systems. Equipped with advanced technology that provides flexibility, sensitivity, and convenience for a wide range of assays such as ELISA, microbial growth, and protein quantitation, all in an incredibly small footprint.

  • Compact design
  • Minimum volume and maximum throughput with 96-well microplate compatibility
  • Powerful data analysis with SoftMax® Pro Software
  • Optimal performance with validation tools

SpectraMax® ABS readers use a grating monochromator to select the exact wavelength needed for every assay. The ABS reader ranges from 340 to 850 nm in wavelength. The advanced optical and electronic design of the readers give the same high performance with round-bottom, flat-bottom or half-area well plates. Select upto six wavelengths in a single read to maximise the number of data points collected in an experiment.

 

 

  • Catalog No:
  • 736-0916
  • MLDVABS-ACAD3
  • Model:
  • SpectraMax® ABS
  • SpectraMax® ABS
  • Lyskilde:
  • Xenon flash lamp
  • Xenon flash lamp
  • Måletilstande:
  • End-point and kinetics
  • End-point and kinetics
  • Bølgelængdeområde:
  • 340 - 850 nm
  • 340 - 850 nm
  • Valg af bølgelængde:
  • Monochromator, tunable in 1,0 nm increments
  • Monochromator, tunable in 1,0 nm increments
  • Bølgelængde nøjagtighed:
  • ±1 nm
  • ±1 nm
  • Wavelength bandwidth:
  • 2,0 nm
  • 2,0 nm
  • Gentagelsesnøjagtighed for bølgelængde:
  • ±0,2 nm
  • ±0,2 nm
  • Inkubationsområde:
  • 5 °C above ambient to 45 °C
  • 5 °C above ambient to 45 °C
  • Måletid:
  • 9 s in normal read mode, 5 s in speed read mode
  • 9 s in normal read mode, 5 s in speed read mode
  • Fotometrisk nøjagtighed:
  • <±0,006 OD ±1%
  • <±0,006 OD ±1%
  • Photometric linearity:
  • 0,0 - 3,0 OD
  • 0,0 - 3,0 OD
  • Fotometrisk område:
  • 0,000 - 4,000 OD
  • 0,000 - 4,000 OD
  • Pladetyper:
  • 96-well plates
  • 96-well plates
  • Rystning:
  • Linear
  • Linear
  • Software:
  • SoftMax® Pro
  • SoftMax® Pro
  • Straylight:
  • <0,05 % at 230 nm
  • <0,05 % at 230 nm
  • Temperature uniformity:
  • ±0,5 °C
  • ±0,5 °C
  • B×D×H:
  • 305×414×203 mm
  • 305×414×203 mm
  • Vægt:
  • 11,3 kg
  • 11,3 kg

 

Product Family Options

 

  • Beskrivelse
     
     
  • SpectraMax® ABS microplate reader, single-mode, with current version of SoftMax® Pro
       
       
    1 * 1 Items
    1 * 1 Items

     
     
    Specifications
    Cat. No.
    736-0916
    Model
    SpectraMax® ABS
    Lyskilde
    Xenon flash lamp
    Måletilstande
    End-point and kinetics
    Bølgelængdeområde
    340 - 850 nm
    Valg af bølgelængde
    Monochromator, tunable in 1,0 nm increments
    Bølgelængde nøjagtighed
    ±1 nm
    Wavelength bandwidth
    2,0 nm
    Gentagelsesnøjagtighed for bølgelængde
    ±0,2 nm
    Inkubationsområde
    5 °C above ambient to 45 °C
    Måletid
    9 s in normal read mode, 5 s in speed read mode
    Fotometrisk nøjagtighed
    <±0,006 OD ±1%
    Photometric linearity
    0,0 - 3,0 OD
    Fotometrisk område
    0,000 - 4,000 OD
    Pladetyper
    96-well plates
    Rystning
    Linear
    Software
    SoftMax® Pro
    Straylight
    <0,05 % at 230 nm
    Temperature uniformity
    ±0,5 °C
    B×D×H
    305×414×203 mm
    Vægt
    11,3 kg
  • SpectraMax® ABS microplate reader, single-mode, with current version of SoftMax® Pro and 2 additional years MD service plan covering parts and labour
       
       
    1 * 1 Items
    1 * 1 Items

     
     
    Specifications
    Cat. No.
    MLDVABS-ACAD3
    Model
    SpectraMax® ABS
    Lyskilde
    Xenon flash lamp
    Måletilstande
    End-point and kinetics
    Bølgelængdeområde
    340 - 850 nm
    Valg af bølgelængde
    Monochromator, tunable in 1,0 nm increments
    Bølgelængde nøjagtighed
    ±1 nm
    Wavelength bandwidth
    2,0 nm
    Gentagelsesnøjagtighed for bølgelængde
    ±0,2 nm
    Inkubationsområde
    5 °C above ambient to 45 °C
    Måletid
    9 s in normal read mode, 5 s in speed read mode
    Fotometrisk nøjagtighed
    <±0,006 OD ±1%
    Photometric linearity
    0,0 - 3,0 OD
    Fotometrisk område
    0,000 - 4,000 OD
    Pladetyper
    96-well plates
    Rystning
    Linear
    Software
    SoftMax® Pro
    Straylight
    <0,05 % at 230 nm
    Temperature uniformity
    ±0,5 °C
    B×D×H
    305×414×203 mm
    Vægt
    11,3 kg