Active pirani gauges
Leverandør: EDWARDS HIGH VACUUM METTLER
Om denne vare
Model APG.
These gauges have a measuring range from atmosphere to 1×10⁻⁴ mbar and combine the gauge head and controller in one compact active unit.
- Wide range, regulated, internal power supply - runs from standard DC power supplies 20 to 36 V
- Tolerant to voltage fluctuations
- Medium and low-pressure versions available - APG-M and -MP from atmosphere to 10⁻³ mbar. APG-L from atmosphere to 10⁻⁴ mbar
- Standard analogue output 0 to 10 V - Easy to interface with a computer or plc. Fault output indication
- Adjustable set-point for process control and interlocking
- Provides vacuum status indication
- Set-point level read visually or with voltmeter
2 tilgængelige muligheder nedenfor
Product Details & Documents
Model APG.
These gauges have a measuring range from atmosphere to 1×10⁻⁴ mbar and combine the gauge head and controller in one compact active unit.
- Wide range, regulated, internal power supply - runs from standard DC power supplies 20 to 36 V
- Tolerant to voltage fluctuations
- Medium and low-pressure versions available - APG-M and -MP from atmosphere to 10⁻³ mbar. APG-L from atmosphere to 10⁻⁴ mbar
- Standard analogue output 0 to 10 V - Easy to interface with a computer or plc. Fault output indication
- Adjustable set-point for process control and interlocking
- Provides vacuum status indication
- Set-point level read visually or with voltmeter
Specifications for Product Family
Specifikationer
- Catalog No:
- 619-0220
- 619-0221
- Model:
- APG-M-NW16
- APG-L-NW16
- Nøjagtighed:
- 1×10ˉ3 mbar
- 1×10ˉ4 mbar
- Tilladt driftstemperatur:
- +10...+60 °C
- +10...+60 °C
- Tilladt rumtemperatur under opbevaring:
- 0...+70 °C
- 0...+70 °C
- Effekt, W:
- 20 to 36 V dc (max 1 V ripple)
- 20 to 36 V dc (max 1 V ripple)
Specifikationer
Product Family Options
Produktinformation
- TilgængelighedPrice
- 619-0220Active pirani gauges, APG-M-NW16Specifikationer:
More
Specifications
Cat. No.619-0220ModelAPG-M-NW16Nøjagtighed1×10ˉ3 mbarTilladt driftstemperatur+10...+60 °CTilladt rumtemperatur under opbevaring0...+70 °CEffekt, W20 to 36 V dc (max 1 V ripple) - 619-0221Active pirani gauges, APG-L-NW16Specifikationer:
More
Specifications