Flashpoint testers, Setaflash 3
Four versions of the closed cup flashpoint testers are available. All functions are accessed by a two-button keypad and the SMART control feature (Seta Multifunctional And Rotational Test). The digital display shows test parameters, instrument status and the test results. Audible prompts are sounded when an action is required by the user. The automatic flash detector reduces the chance of misinterpretation of the results and enhances repeatability. Instruments have a integral gas tank which is filled using a standard butane (lighter) refill cartridge and are suitable for mobile use when operated with the optional 12 volt power supply adapter ( please ask for details).
- Flash point tests in less than two minutes
- Small sample sizes of 2 or 4 ml
- Audible and on screen prompts for ease of use
Closed cup tester: As above with the addition of a temperature ramp mode that allow the rapid determination of the flash point of a sample in accordance with ASTM D7236 and IP 534.
Active cool closed cup tester: Active Cool function to rapidly and efficiently carry out Flash/No flash tests and extend the temperature down to 10 °C - Cooling and heating with Peltier cells.
Active cool corrosion resisting closed cup tester: Flash/No Flash testing (10 to 130 °C) rapid temperature normalisation between tests. Corrosion resisting insert in the sample cup.
: Basic models: ASTM D3278, BS EN456, ISO 3680, UN class 3 non viscous flammable liquids, ASTM E502, ISO 3679, BS 3900 Part A13, EN 456, ASTM D3828, IP 303-1/2, CHIPS regulations, classification of dangerous goods for carriage: Viscous and non viscous liquids, EPA 1020 A/B, IP 523, 524.
Models with ramp mode. As above plus ASTM D7236 and IP 534.
: Supplied with a syringe, mains lead and instruction manual.
Four versions of the closed cup flashpoint testers are available. All functions are accessed by a two-button keypad and the SMART control feature (Seta Multifunctional And Rotational Test). The digital display shows test parameters, instrument status and the test results. Audible prompts are sounded when an action is required by the user. The automatic flash detector reduces the chance of misinterpretation of the results and enhances repeatability. Instruments have a integral gas tank which is filled using a standard butane (lighter) refill cartridge and are suitable for mobile use when operated with the optional 12 volt power supply adapter ( please ask for details).
- Flash point tests in less than two minutes
- Small sample sizes of 2 or 4 ml
- Audible and on screen prompts for ease of use
Closed cup tester: As above with the addition of a temperature ramp mode that allow the rapid determination of the flash point of a sample in accordance with ASTM D7236 and IP 534.
Active cool closed cup tester: Active Cool function to rapidly and efficiently carry out Flash/No flash tests and extend the temperature down to 10 °C - Cooling and heating with Peltier cells.
Active cool corrosion resisting closed cup tester: Flash/No Flash testing (10 to 130 °C) rapid temperature normalisation between tests. Corrosion resisting insert in the sample cup.
: Basic models: ASTM D3278, BS EN456, ISO 3680, UN class 3 non viscous flammable liquids, ASTM E502, ISO 3679, BS 3900 Part A13, EN 456, ASTM D3828, IP 303-1/2, CHIPS regulations, classification of dangerous goods for carriage: Viscous and non viscous liquids, EPA 1020 A/B, IP 523, 524.
Models with ramp mode. As above plus ASTM D7236 and IP 534.
: Supplied with a syringe, mains lead and instruction manual.
- Catalog No:
- 751-0013
- 751-0065
- 751-0066
- 751-0067
- Model:
- 31000-0
- 33000-0
- 33200-2
- 33250-2
- Temperature range:
- Ambient…300 °C
- Ambient…300 °C
- 10…130 °C
- 10…130 °C
- Test modes:
- Rapid equilibrium and ramp 2 °C/minute
- Rapid equilibrium and ramp 2 °C/minute
- Rapid equilibrium and ramp 2 °C/minute
- Rapid equilibrium and ramp 2 °C/minute
- Cup material:
- Blackened aluminium
- Aluminium
- Aluminium
- Corrosion resistant steel
- Sample size:
- 2 ml for flash points up to 100 °C; 4 ml for flash points from 100 to 300 °C
- 2 ml for flash points up to 100 °C; 4 ml for flash points from 100 to 300 °C
- 2 ml for flash points up to 100 °C; 4 ml for flash points from 100 to 300 °C
- 2 ml for flash points up to 100 °C; 4 ml for flash points from 100 to 300 °C
- Test time:
- 1 to 99 min.
- Test duration RE mode:
- 1 min for flash points up to 100 °C, 2 min for flash points 100 to 300 °C or user determined 1 to 99 min
- 1 min for flash points up to 100 °C, 2 min for flash points 100 to 300 °C or user determined 1 - 99 min
- 1 min for flash points up to 100 °C, 2 min for flash points 100 to 300 °C or user determined 1 - 99 min
- 1 min for flash points up to 100 °C, 2 min for flash points 100 to 300 °C or user determined 1 - 99 min
- Power:
- 200 W
- 200 W
- 200 W
- 200 W
- W×D×H:
- 280×256×256
- 280×260×260 mm
- 280×260×260 mm
- 280×260×260 mm
- Weight:
- 4 kg
- 4 kg
- 5 kg
- 5 kg
Product Family Options
- DescriptionModel
Specifications
Cat. No.751-0013Model31000-0Temperature rangeAmbient…300 °CTest modesRapid equilibrium and ramp 2 °C/minuteCup materialBlackened aluminiumSample size2 ml for flash points up to 100 °C; 4 ml for flash points from 100 to 300 °CTest time1 to 99 min.Test duration RE mode1 min for flash points up to 100 °C, 2 min for flash points 100 to 300 °C or user determined 1 to 99 minPower200 WW×D×H280×256×256Weight4 kgSpecifications
Cat. No.751-0065Model33000-0Temperature rangeAmbient…300 °CTest modesRapid equilibrium and ramp 2 °C/minuteCup materialAluminiumSample size2 ml for flash points up to 100 °C; 4 ml for flash points from 100 to 300 °CTest timeTest duration RE mode1 min for flash points up to 100 °C, 2 min for flash points 100 to 300 °C or user determined 1 - 99 minPower200 WW×D×H280×260×260 mmWeight4 kgSpecifications
Cat. No.751-0066Model33200-2Temperature range10…130 °CTest modesRapid equilibrium and ramp 2 °C/minuteCup materialAluminiumSample size2 ml for flash points up to 100 °C; 4 ml for flash points from 100 to 300 °CTest timeTest duration RE mode1 min for flash points up to 100 °C, 2 min for flash points 100 to 300 °C or user determined 1 - 99 minPower200 WW×D×H280×260×260 mmWeight5 kgSpecifications
Cat. No.751-0067Model33250-2Temperature range10…130 °CTest modesRapid equilibrium and ramp 2 °C/minuteCup materialCorrosion resistant steelSample size2 ml for flash points up to 100 °C; 4 ml for flash points from 100 to 300 °CTest timeTest duration RE mode1 min for flash points up to 100 °C, 2 min for flash points 100 to 300 °C or user determined 1 - 99 minPower200 WW×D×H280×260×260 mmWeight5 kg
Certified thermometers