Multi-parameter meter (pH/mV/conductivity), bench, Orion Star™ A215
About this item
This benchtop meter is ideal for a wide range of applications and advanced pH and conductivity analysis in the lab. It offers two measuring channels allowing users to measure pH and conductivity simultaneously or view each channel separately.
- Informative, easy-to-read backlit graphic display with clear onscreen instructions simplifies operation
- Measurement stability indicator and selectable read modes – Auto-Read holds a stable reading, Continuous shows changing readings with hold option and Timed logs data at specific intervals
- Data log collects and date/time stamps up to 2000 measurement sets with optional sample ID and user ID
- Calibration log stores ten most recent calibrations
- Non volatile meter memory ensures your data and settings are protected, even with loss of power
- Orion Star Com software facilitates data transfer to a computer via USB or RS232 connections
- Flexible power options with universal power adapter or optional AA batteries
IP54-rated housing resists dust and splashes and can be wall-mounted if table space is limited.
Measure pH, mV, relative mV or ORP with temperature on channel one and measure conductivity, TDS, salinity or resistivity with temperature on channel two.
Perform up to a five point pH calibration using automatic buffer recognition of NIST or DIN buffer groups with choice to manually enter custom buffer values. Perform up to a five point conductivity calibration. Calibration editing option allows individual points to be fixed without a full recalibration. Auto-ranging conductivity values offer consistent readings across entire measurement range without need to recalibrate.
Use linear, non linear, EP (USP) or off conductivity temperature compensation with 5, 10, 15, 20 or 25 °C reference temperatures.
Certifications: CE, TUV 3-in-1 and FCC Class A certifications.
Delivery information: Supplied with electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable, meter test certificate and 3-year replacement warranty.
The kit additionally includes: 8157BNUMD ROSS Ultra Triode refillable epoxy-body pH/ATC electrode, 013005MD DuraProbe 4-cell (K=0,475) epoxy-body conductivity sensor, ROSS pH buffer and storage solution kit and Orion 1413 µS conductivity standard (5× 60 ml).
2 Options Available Below
Product Details & Documents
This benchtop meter is ideal for a wide range of applications and advanced pH and conductivity analysis in the lab. It offers two measuring channels allowing users to measure pH and conductivity simultaneously or view each channel separately.
- Informative, easy-to-read backlit graphic display with clear onscreen instructions simplifies operation
- Measurement stability indicator and selectable read modes – Auto-Read holds a stable reading, Continuous shows changing readings with hold option and Timed logs data at specific intervals
- Data log collects and date/time stamps up to 2000 measurement sets with optional sample ID and user ID
- Calibration log stores ten most recent calibrations
- Non volatile meter memory ensures your data and settings are protected, even with loss of power
- Orion Star Com software facilitates data transfer to a computer via USB or RS232 connections
- Flexible power options with universal power adapter or optional AA batteries
IP54-rated housing resists dust and splashes and can be wall-mounted if table space is limited.
Measure pH, mV, relative mV or ORP with temperature on channel one and measure conductivity, TDS, salinity or resistivity with temperature on channel two.
Perform up to a five point pH calibration using automatic buffer recognition of NIST or DIN buffer groups with choice to manually enter custom buffer values. Perform up to a five point conductivity calibration. Calibration editing option allows individual points to be fixed without a full recalibration. Auto-ranging conductivity values offer consistent readings across entire measurement range without need to recalibrate.
Use linear, non linear, EP (USP) or off conductivity temperature compensation with 5, 10, 15, 20 or 25 °C reference temperatures.
Certifications: CE, TUV 3-in-1 and FCC Class A certifications.
Delivery information: Supplied with electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable, meter test certificate and 3-year replacement warranty.
The kit additionally includes: 8157BNUMD ROSS Ultra Triode refillable epoxy-body pH/ATC electrode, 013005MD DuraProbe 4-cell (K=0,475) epoxy-body conductivity sensor, ROSS pH buffer and storage solution kit and Orion 1413 µS conductivity standard (5× 60 ml).
Specifications for Product Family
Specifications
- Catalog No:
- 665-0272
- 665-0271
- Model:
- A215
- A215
- pH range:
- –2,000...+20,000
- –2,000...+20,000
- pH resolution:
- 0,1 / 0,01 / 0,001
- 0,1 / 0,01 / 0,001
- pH accuracy:
- ±0,002
- ±0,002
- pH calibration:
- Up to 5 points; Automatic buffer recognition (USA/NIST and DIN)
- Up to 5 points; Automatic buffer recognition (USA/NIST and DIN)
- mV range:
- ±2000,0
- ±2000,0
- mV resolution:
- 0,1
- 0,1
- mV accuracy:
- ±0,2 mV or ±0,05% of reading, whichever is greater
- ±0,2 mV or ±0,05% of reading, whichever is greater
- Conductivity range:
- 0,001 µS/cm - 3000 mS/cm
- 0,001 µS/cm - 3000 mS/cm
- Conductivity resolution:
- 0,001 μS minimum, auto ranging, up to 4 significant digits
- 0,001 μS minimum, auto ranging, up to 4 significant digits
- Conductivity accuracy:
- 0,5% of reading ±1 digit > 3 μS, 0,5% of reading ±0,01 μS ≤ 3 μS
- 0,5% of reading ±1 digit > 3 μS, 0,5% of reading ±0,01 μS ≤ 3 μS
- Conductivity cell constant:
- 0,001 - 199,9 cm⁻¹
- 0,001 - 199,9 cm⁻¹
- Conductivity calibration:
- Up to 5 points
- Up to 5 points
- Resistivity range:
- 2 Ω/cm - 100 MΩ/cm
- 2 Ω/cm - 100 MΩ/cm
- Resistivity accuracy:
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- Resistivity resolution:
- 1 ohm or 0,1 meg-ohm, auto ranging
- 1 ohm or 0,1 meg-ohm, auto ranging
- Salinity range:
- 0,06 - 80,00 psu / 0,05 - 42,00 ppt
- 0,06 - 80,00 psu / 0,05 - 42,00 ppt
- Salinity resolution:
- 0,01 psu or 0,01 ppt minimum, auto ranging
- 0,01 psu or 0,01 ppt minimum, auto ranging
- Salinity accuracy:
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- TDS range:
- 0,001 - 200,0 ppm
- 0,001 - 200,0 ppm
- TDS resolution:
- 0,001 ppm minimum, auto ranging, up to 4 significant digits
- 0,001 ppm minimum, auto ranging, up to 4 significant digits
- TDS accuracy:
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- Temperature range:
- –5,0...+105,0 °C
- –5,0...+105,0 °C
- Temperature resolution:
- 0,1 °C
- 0,1 °C
- Temperature accuracy:
- ±0,1 °C
- ±0,1 °C
- Temperature compensation:
- Linear, non linear nLFn, non linear nLFu, EP (USP), off
- Linear, non linear nLFn, non linear nLFu, EP (USP), off
- Sensor type:
- DuraProbe 4-cell conductivity sensor
- DuraProbe 4-cell conductivity sensor
- Display:
- Graphic LCD with backlight
- Graphic LCD with backlight
- Battery life:
- 800 h
- 800 h
- Power supply:
- AC adapter or 4 AA batteries
- AC adapter or 4 AA batteries
- Compliance:
- IP 54
- IP 54
- Operating temperature range:
- 5...45 °C (ambient)
- 5...45 °C (ambient)
- Data logging - log function:
- Manual / Automatic with AUTO-READ™
- Manual / Automatic with AUTO-READ™
- Data logging capacity:
- 2000 with date and time stamp
- 2000 with date and time stamp
- IP rating:
- IP 54
- IP 54
- Connections:
- BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)
- BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)
- Outputs:
- RS232, USB
- RS232, USB
- Input:
- BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)
- BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)
- Reference temperature:
- 5 / 10 / 15 / 20 / 25 °C
- 5 / 10 / 15 / 20 / 25 °C
- Alarm:
- High/low limit alarm, calibration due alarm
- High/low limit alarm, calibration due alarm
- No. of channels:
- 2
- 2
- Calibration:
- pH with calibration editing option, relative mV (RmV), ORP, conductivity with calibration editing option
- pH with calibration editing option, relative mV (RmV), ORP, conductivity with calibration editing option
- Calibration points:
- 1 point relative mV (RmV) mode, 1 point ORP mode
- 1 point relative mV (RmV) mode, 1 point ORP mode
- W×D×H:
- 240×180×110 mm
- 240×180×110 mm
- Weight:
- 0,9 kg
- 0,9 kg
Specifications
Product Family Options
Product Information
- AvailabilityPrice
- 665-0272Bench pH/conductivity meter Orion Star A215, kit incl. ROSS Ultra Triode pH/ATC probe and DuraProbe conductivity cellBench pH/conductivity meter Orion Star A215, kit incl. ROSS Ultra Triode pH/ATC probe and DuraProbe conductivity cellLog in for availability1 * 1 ST2 320,00 €Specifications:
More
Specifications
Cat. No.665-0272ModelA215pH range–2,000...+20,000pH resolution0,1 / 0,01 / 0,001pH accuracy±0,002pH calibrationUp to 5 points; Automatic buffer recognition (USA/NIST and DIN)mV range±2000,0mV resolution0,1mV accuracy±0,2 mV or ±0,05% of reading, whichever is greaterConductivity range0,001 µS/cm - 3000 mS/cmConductivity resolution0,001 μS minimum, auto ranging, up to 4 significant digitsConductivity accuracy0,5% of reading ±1 digit > 3 μS, 0,5% of reading ±0,01 μS ≤ 3 μSConductivity cell constant0,001 - 199,9 cm⁻¹Conductivity calibrationUp to 5 pointsResistivity range2 Ω/cm - 100 MΩ/cmResistivity accuracy0,5% of reading ±1 digitResistivity resolution1 ohm or 0,1 meg-ohm, auto rangingSalinity range0,06 - 80,00 psu / 0,05 - 42,00 pptSalinity resolution0,01 psu or 0,01 ppt minimum, auto rangingSalinity accuracy0,5% of reading ±1 digitTDS range0,001 - 200,0 ppmTDS resolution0,001 ppm minimum, auto ranging, up to 4 significant digitsTDS accuracy0,5% of reading ±1 digitTemperature range–5,0...+105,0 °CTemperature resolution0,1 °CTemperature accuracy±0,1 °CTemperature compensationLinear, non linear nLFn, non linear nLFu, EP (USP), offSensor typeDuraProbe 4-cell conductivity sensorDisplayGraphic LCD with backlightBattery life800 hPower supplyAC adapter or 4 AA batteriesComplianceIP 54Operating temperature range5...45 °C (ambient)Data logging - log functionManual / Automatic with AUTO-READ™Data logging capacity2000 with date and time stampIP ratingIP 54ConnectionsBNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)OutputsRS232, USBInputBNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)Reference temperature5 / 10 / 15 / 20 / 25 °CAlarmHigh/low limit alarm, calibration due alarmNo. of channels2CalibrationpH with calibration editing option, relative mV (RmV), ORP, conductivity with calibration editing optionCalibration points1 point relative mV (RmV) mode, 1 point ORP modeW×D×H240×180×110 mmWeight0,9 kg - 665-0271Bench pH/conductivity meter Orion Star A215, instrument aloneBench pH/conductivity meter Orion Star A215, instrument aloneLog in for availability1 * 1 ST1 480,00 €Specifications:
More
Specifications
Cat. No.665-0271ModelA215pH range–2,000...+20,000pH resolution0,1 / 0,01 / 0,001pH accuracy±0,002pH calibrationUp to 5 points; Automatic buffer recognition (USA/NIST and DIN)mV range±2000,0mV resolution0,1mV accuracy±0,2 mV or ±0,05% of reading, whichever is greaterConductivity range0,001 µS/cm - 3000 mS/cmConductivity resolution0,001 μS minimum, auto ranging, up to 4 significant digitsConductivity accuracy0,5% of reading ±1 digit > 3 μS, 0,5% of reading ±0,01 μS ≤ 3 μSConductivity cell constant0,001 - 199,9 cm⁻¹Conductivity calibrationUp to 5 pointsResistivity range2 Ω/cm - 100 MΩ/cmResistivity accuracy0,5% of reading ±1 digitResistivity resolution1 ohm or 0,1 meg-ohm, auto rangingSalinity range0,06 - 80,00 psu / 0,05 - 42,00 pptSalinity resolution0,01 psu or 0,01 ppt minimum, auto rangingSalinity accuracy0,5% of reading ±1 digitTDS range0,001 - 200,0 ppmTDS resolution0,001 ppm minimum, auto ranging, up to 4 significant digitsTDS accuracy0,5% of reading ±1 digitTemperature range–5,0...+105,0 °CTemperature resolution0,1 °CTemperature accuracy±0,1 °CTemperature compensationLinear, non linear nLFn, non linear nLFu, EP (USP), offSensor typeDuraProbe 4-cell conductivity sensorDisplayGraphic LCD with backlightBattery life800 hPower supplyAC adapter or 4 AA batteriesComplianceIP 54Operating temperature range5...45 °C (ambient)Data logging - log functionManual / Automatic with AUTO-READ™Data logging capacity2000 with date and time stampIP ratingIP 54ConnectionsBNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)OutputsRS232, USBInputBNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)Reference temperature5 / 10 / 15 / 20 / 25 °CAlarmHigh/low limit alarm, calibration due alarmNo. of channels2CalibrationpH with calibration editing option, relative mV (RmV), ORP, conductivity with calibration editing optionCalibration points1 point relative mV (RmV) mode, 1 point ORP modeW×D×H240×180×110 mmWeight0,9 kg
Accessories
Product Information
- Catalog #ForAvailabilityPrice
- Printer and computer cable set, incl. one RS232 printer interface cable and one RS232 computer interface cablePrinter and computer cable set, incl. one RS232 printer interface cable and one RS232 computer interface cable662-1140Thermo Scientific™ Orion™ Versa Star meters, Orion™ Star A series meters (Star A320, Star A220 and Star A210 series) and legacy Orion™ Star and Star Plus metersLog in for availability1 * 1 SET - 76,40 €
Product Information
- Catalog #AvailabilityPrice
- Spare universal power adapter for Thermo Scientific™ Orion™ Star A series benchtop and portable meters and legacy Orion™ Star and Star Plus series metersSpare universal power adapter for Thermo Scientific™ Orion™ Star A series benchtop and portable meters and legacy Orion™ Star and Star Plus series meters662-1547Log in for availability1 * 1 ST - 72,40 €
Product Information
- Catalog #AvailabilityPrice
- Orion™ Star™ A Series weighted base for use with meter-attached electrode standOrion™ Star™ A Series weighted base for use with meter-attached electrode standORIOSTARA-HBLog in for availability1 * 1 ST - 56,30 €
Product Information
- Catalog #ForAvailabilityPrice
- Cable adapter, for connecting the RS232 port of a meter to a USB portCable adapter, for connecting the RS232 port of a meter to a USB portORIO1010005Orion™ Versa Star, Orion™ Star A series (A320, A220 + A210 series) and Orion™ Star and Star Plus metersLog in for availability1 * 1 ST - 131,00 €
Product Information
- Catalog #ForAvailabilityPrice
- RS232 cable, for connecting a meter to a PC or printerRS232 cable, for connecting a meter to a PC or printerORIO1010053Orion™ Versa Star, Orion™ Star A series (A320, A220 + A210 series), Orion™ Dual Star benchtop meters, Orion 3-Star, 4-Star and 5-Star metersLog in for availability1 * 1 ST - 45,40 €
Product Information
- Catalog #AvailabilityPrice
- Benchtop meter-attached electrode stand (electrode holder arm and head), Orion™ Star A series, 360° motion (without weighted base)Benchtop meter-attached electrode stand (electrode holder arm and head), Orion™ Star A series, 360° motion (without weighted base)ORIOSTARA-BEALog in for availability1 * 1 ST - 163,00 €
Recommendations will be personalized based on your shopping preferences only if you have given your consent by enabling "Enhance my Shopping Experience" on the "Personal Info page".