Order Entry
China
ContactUsLinkComponent
Orion Star™ A326 Multi-Parameter Meter (pH/RDO/DO), Handheld
Orion Star™ A326 Multi-Parameter Meter (pH/RDO/DO), Handheld
  ORIOSTARA3260
undefined
Orion Star™ A326 Multi-Parameter Meter (pH/RDO/DO), Handheld
  ORIOSTARA3260

 

  • Calibration:
    pH with calibration editing option, relative mV (RmV), ORP, dissolved oxygen and temperature
  • Memory:
    Non-volatile memory preserves data log, calibration log and meter settings
  • No. of channels:
    2
  • Temperature range:
    –5,0...+105,0 °C
  • Description:
    Handheld pH/DO meter Orion Star A326, instrument alone
  • Seq:
    01
  • Model:
    A326
  • Temperature accuracy:
    ±0,1 °C
  • Temperature resolution:
    0,1 °C, 0,1 °F
  • Pressure resolution:
    1 mm Hg
  • Pressure accuracy:
    6 mm Hg
  • Battery life:
    800 hours
  • DO resolution:
    0.1, 1 % saturation
  • IP rating:
    IP 67
  • Alarm:
    High/low limit alarm, calibration due alarm
  • Calibration points:
    1 point relative mV (RmV) mode, 1 point ORP mode
  • Compliance:
    IP 67
  • Connections:
    BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (ATC temperature probe), 9 pin MiniDIN (RDO optical or polarographic DO sensor with built-in temperature)
  • DO barometric pressure range:
    Automatic or manual, 450 - 850 mmHg
  • DO calibration:
    Water-saturated air, air saturated water, manual (Winkler) and zero point
  • DO concentration:
    0 - 90 mg/L (polarographic); 0 - 50 mg/L (RDO)
  • DO concentration resolution:
    0,1 / 0,01 mg/L
  • DO saturation:
    0,0 - 600,0 % (polarographic); 0,0 - 500,0 % (RDO)
  • DO accuracy:
    Polarographic: ±0,2 mg/L or ±2% of reading, whichever is greater
    RDO: ±0,1 mg/L (up to 8 mg/L), ±0,2 mg/L (8 - 20 mg/L), ±10% of reading (up to 50 mg/L)
  • Data logging capacity:
    5000 with date and time stamp
  • Data logging - log function:
    Manual / Automatic with AUTO-READ™
  • Display:
    Graphic LCD with backlight
  • Languages:
    English, Spanish, French, Italian, German and Chinese; Portuguese and Korean available with software update
  • Outputs:
    RS232, USB
  • Power supply:
    AC adapter or 4 AA batteries
  • Temperature calibration:
    1 point temperature offset calibration
  • Temperature compensation:
    Automatic or manual
  • Weight:
    0,5 kg
  • W×D×H:
    240×100×65 mm
  • mV accuracy:
    ±0,2 mV or 0,05% of reading whichever is greater
  • mV range:
    ±2000,0
  • mV resolution:
    0,1
  • pH accuracy:
    ±0,002
  • pH calibration:
    Up to 5 points; Automatic buffer recognition (USA/NIST and DIN)
  • pH range:
    –2,000...+20,000
  • pH resolution:
    0,1 / 0,01 / 0,001
  • Battery type:
    4 AA (included)
  • Pk:
    1

 

 

This rugged, waterproof portable meter is designed for a wide range of pH, mV, ORP, dissolved oxygen and temperature testing and field applications. It can be used in the most demanding locations thanks to its IP67-rated housing. The meter offers two measuring channels allowing users to measure pH and dissolved oxygen simultaneously or view each channel separately.

  • Up to a five point pH calibration using automatic buffer recognition of NIST or DIN buffer groups
  • Automatically compensate dissolved oxygen readings using ATC temperature, manual salinity input and built-in barometer for barometric pressure readings
  • Informative, easy-to-read backlit graphic display with clear onscreen instructions simplifies operation
  • Measurement stability indicator and selectable read modes – Auto-Read holds a stable reading, Continuous shows changing readings with hold option and Timed logs data at specific intervals
  • Data log collects and date/time stamps up to 5000 measurement sets with optional sample ID and user ID
  • Calibration log stores ten most recent calibrations
  • Orion Star Com software facilitates data transfer to a computer via USB or RS232 connections

Measure pH, mV, relative mV or ORP with temperature on channel one and measure dissolved oxygen in percent saturation or mg/L with temperature using RDO optical DO probes or polarographic DO probes on channel two. Warrenty; 3 years.

 : CE, TUV 3-1, FCC Class A.
 : Supplied with four AA batteries, literature CD, printed quick start guide, computer interface cable and meter test certificate.
The kit additionally includes: 8107UWMMD ROSS gel-filled epoxy pH/ATC Triode (3 m cable), 087010MD RDO optical DO probe (3 m cable), pH 4,01/pH 7,00/pH 10,01 buffer pouches, Orion ROSS electrode storage solution, Orion 1413 µS conductivity standard pouches, rinse solution pouches, protective armor with electrode holder and hard carrying case.