Inserisci ordine
Switzerland
Orders LinkContactUsLinkComponent
 

Temperature Data Loggers for Process Validation, EBI 12-T441

 :   EBRO
Temperature Data Loggers for Process Validation, EBI 12-T441
 :  EBRO
 
Temperature Data Loggers for Process Validation, EBI 12-T441
Temperature Data Loggers for Process Validation, EBI 12-T441

 

 

The EBI 12-T441 is a precise temperature data logger for process validation in laboratories and the pharmaceutical industry. With a temperature range of −200 to +200 °C and an accuracy of up to 0,05 °C, it enables precise temperature measurements in critical applications such as sterilisation, disinfection and ovens.

  •  Suitable for steam sterilisers, washer-disinfectors

  • Temperature range −200 to +200 °C
  • High accuracy up to 0,05 °C
  • Two bendable metal probe (Ø1,50 mm)

The EBI 12-T441 temperature data logger from ebro fulfils the stringent requirements of the pharmaceutical industry and laboratory applications. It is ideal for process validation in steam sterilisers, washer-disinfectors and other critical processes where precise temperature control is required. With a wide temperature range from −200 to +200 °C and an accuracy of up to 0,05 °C, it ensures reliable results. Thanks to real-time data transmission and flexible measurement interval selection, it is perfect for seamless documentation in validated processes. Its robust design ensures a long service life and high resistance in demanding environments.

Dimensions (12-T441/12-T441-EX): Ø48×24H mm, probe length 500 mm
Operating temperature (12-T441/12-T441-EX): −90…+150 °C
Calibration points (12-T441/12-T441-EX): −20 °C, 0 °C, +60 °C and +134 °C

 : Certified according to ISO 17665, ISO 15883, (EC) 1935/2004​.

 : 1× EBI 12-T441, 1× factory calibration certificate, manual.

 

 

  • Catalog No:
  • EBRO1340-6629-EX
  • EBRO1340-6629
  • Modello:
  • EBI 12-T441-EX
  • EBI 12-T441
  • Sensore:
  • PT1000, class A; Axial, 2 external temperature probe, bendable, Ø 1,5 mm, 500 mm
  • PT1000, class A; Axial, 2 external temperature probe, bendable, Ø 1,5 mm, 500 mm
  • Classe di protezione:
  • IP 68
  • IP 68
  • Intervalli di registrazione:
  • 250 ms … 24 hrs.
  • 250 ms … 24 hrs.
  • Peso:
  • 90 g
  • 90 g
  • Memoria dati - N. letture:
  • 2×50.160
  • 2×50.160
  • Batteria:
  • Lithium, 2× BR1225A, 3 V user–replaceable
  • Lithium, 2× BR1225A, 3 V user–replaceable

 

Product Family Options

01_Loggers

 

  • Descrizione
    Range di misura
    Accuratezza
    Risoluzione
     
     
  • EBI 12-T441-EX high temperature data logger with two bendable probes; L= 500 mm; permissible ambient conditions in the hazardous area: −40…+85 °C, 100 Pa…400 kPa; ATEX-type: II1G Ex ia IIC T4 Ga
    T: –200…+200 °C
    ±1,5 °C (–200...–90 °C); ±0,5 °C (–90...–40 °C); ±0,2 °C (–40...0 °C); ±0,1 °C (0...+120 °C); ±0,05 °C (120...140 °C); ±0,1 °C (140...150 °C); ±0,5 °C (150...200 °C)
    0,01 °C
       
       
    1 * 1 ST
    1 * 1 ST - SFr. 5’500.00
    SFr. 5’500.00
    1 * 1 ST
       SFr. 5’500.00

     
     
    Specifications
    Cat. No.
    EBRO1340-6629-EX
    Modello
    EBI 12-T441-EX
    Sensore
    PT1000, class A; Axial, 2 external temperature probe, bendable, Ø 1,5 mm, 500 mm
    Classe di protezione
    IP 68
    Intervalli di registrazione
    250 ms … 24 hrs.
    Peso
    90 g
    Memoria dati - N. letture
    2×50.160
    Batteria
    Lithium, 2× BR1225A, 3 V user–replaceable
  • EBI 12-T441 high temperature data logger with two bendable probes, L= 500 mm
    T: –200…+200 °C
    ±1,5 °C (–200...–90 °C); ±0,5 °C (–90...–40 °C); ±0,2 °C (–40...0 °C); ±0,1 °C (0...+120 °C); ±0,05 °C (120...140 °C); ±0,1 °C (140...150 °C); ±0,5 °C (150...200 °C)
    0,01 °C
       
       
    1 * 1 ST
    1 * 1 ST - SFr. 4’140.00
    SFr. 4’140.00
    1 * 1 ST
       SFr. 4’140.00

     
     
    Specifications
    Cat. No.
    EBRO1340-6629
    Modello
    EBI 12-T441
    Sensore
    PT1000, class A; Axial, 2 external temperature probe, bendable, Ø 1,5 mm, 500 mm
    Classe di protezione
    IP 68
    Intervalli di registrazione
    250 ms … 24 hrs.
    Peso
    90 g
    Memoria dati - N. letture
    2×50.160
    Batteria
    Lithium, 2× BR1225A, 3 V user–replaceable

02_Software and readout station

 

  •  
    Descrizione
     
     
  • SI 1200, 4-port Interface IF 200, Software Winlog.pro
       
       
    1 * 1 ST
    1 * 1 ST - SFr. 3’310.00
    SFr. 3’310.00
    1 * 1 ST
       SFr. 3’310.00

 

  •  
    Descrizione
     
     
  • SK 6200, starter kit, food and pharmaceutical application, software Winlog.pro, 4-port interface IF 200, EBI 12-T237, –40…+140 °C
       
       
    1 * 1 ST
    1 * 1 ST - SFr. 2’310.00
    SFr. 2’310.00
    1 * 1 ST
       SFr. 2’310.00

Complementary Products

 

  •  
    Descrizione
    Range di misura
    Risoluzione
    Accuratezza
     
     
  • EBRO1340-6658
    EBI 12-TP237, temperature and pressure data logger, calibration points 0 °C, +60 °C and +134 °C; 100 mbar and 3100 mbar (bei +25 °C); 3100 mbar (at +134 °C)
    T: 0…140 °C
    P: 1…4000 mbar
    0,01 °C
    1 mbar
    ±0,1 (0…140 °C)
    ±20 mbar
       
    1 * 1 ST - SFr. 2’850.00

  • EBRO1340-6600
    EBI 12-T100, high temperature data logger, pharmaceutical, medical and food applications; calibration points: –20, 0, 60 and 134 °C
    T: –90…+150 °C
    0,01 °C
    ±0,5 °C (–90…–40 °C); ±0,2 °C (–40…0 °C); ±0,1 °C (0…120 °C); ±0,05 °C (120…140 °C); ±0,1 °C (140…150 °C)
       
    1 * 1 ST - SFr. 2’340.00

  • EBRO1340-6601-0100
    EBI 12-T101, data logger for washer and disinfecter; calibration points –20 °C, 0 °C and +60 °C
    T: –90…+105 °C
    0,01 °C
    ±0,3 °C (0…105 °C)
    ±0,5 °C (–90 …–40 °C)
       
    1 * 1 ST - SFr. 1’930.00

Accessories

 

  •  
    Descrizione
     
     
  • EBRO1100-0130
    Battery replacement set EBI 12; 3× battery, 6× O-ring, 1× unscrewing aid EBI 12
       
    1 * 1 ST - 322.00 SFr.

 

  •  
    Descrizione
     
     
  • 620-1693
    Battery replacement set EBI 12; 3× battery, 6× O-ring
       
    1 * 1 ST - 226.00 SFr.

Recommendations

Recommendations will be personalized based on your shopping preferences only if you have given your consent by enabling "Enhance my Shopping Experience" on the "Personal Info page".

Otherwise, you will receive generic recommendations.