Multi-parameter meter (pH/mV/conductivity), bench, Orion Star™ A215
À propos de cet article
This benchtop meter is ideal for a wide range of applications and advanced pH and conductivity analysis in the lab. It offers two measuring channels allowing users to measure pH and conductivity simultaneously or view each channel separately.
- Informative, easy-to-read backlit graphic display with clear onscreen instructions simplifies operation
- Measurement stability indicator and selectable read modes – Auto-Read holds a stable reading, Continuous shows changing readings with hold option and Timed logs data at specific intervals
- Data log collects and date/time stamps up to 2000 measurement sets with optional sample ID and user ID
- Calibration log stores ten most recent calibrations
- Non volatile meter memory ensures your data and settings are protected, even with loss of power
- Orion Star Com software facilitates data transfer to a computer via USB or RS232 connections
- Flexible power options with universal power adapter or optional AA batteries
IP54-rated housing resists dust and splashes and can be wall-mounted if table space is limited.
Measure pH, mV, relative mV or ORP with temperature on channel one and measure conductivity, TDS, salinity or resistivity with temperature on channel two.
Perform up to a five point pH calibration using automatic buffer recognition of NIST or DIN buffer groups with choice to manually enter custom buffer values. Perform up to a five point conductivity calibration. Calibration editing option allows individual points to be fixed without a full recalibration. Auto-ranging conductivity values offer consistent readings across entire measurement range without need to recalibrate.
Use linear, non linear, EP (USP) or off conductivity temperature compensation with 5, 10, 15, 20 or 25 °C reference temperatures.
Certifications: CE, TUV 3-in-1 and FCC Class A certifications.
Informations de livraison: Supplied with electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable, meter test certificate and 3-year replacement warranty.
The kit additionally includes: 8157BNUMD ROSS Ultra Triode refillable epoxy-body pH/ATC electrode, 013005MD DuraProbe 4-cell (K=0,475) epoxy-body conductivity sensor, ROSS pH buffer and storage solution kit and Orion 1413 µS conductivity standard (5× 60 ml).
2 variantes
Product Details & Documents
This benchtop meter is ideal for a wide range of applications and advanced pH and conductivity analysis in the lab. It offers two measuring channels allowing users to measure pH and conductivity simultaneously or view each channel separately.
- Informative, easy-to-read backlit graphic display with clear onscreen instructions simplifies operation
- Measurement stability indicator and selectable read modes – Auto-Read holds a stable reading, Continuous shows changing readings with hold option and Timed logs data at specific intervals
- Data log collects and date/time stamps up to 2000 measurement sets with optional sample ID and user ID
- Calibration log stores ten most recent calibrations
- Non volatile meter memory ensures your data and settings are protected, even with loss of power
- Orion Star Com software facilitates data transfer to a computer via USB or RS232 connections
- Flexible power options with universal power adapter or optional AA batteries
IP54-rated housing resists dust and splashes and can be wall-mounted if table space is limited.
Measure pH, mV, relative mV or ORP with temperature on channel one and measure conductivity, TDS, salinity or resistivity with temperature on channel two.
Perform up to a five point pH calibration using automatic buffer recognition of NIST or DIN buffer groups with choice to manually enter custom buffer values. Perform up to a five point conductivity calibration. Calibration editing option allows individual points to be fixed without a full recalibration. Auto-ranging conductivity values offer consistent readings across entire measurement range without need to recalibrate.
Use linear, non linear, EP (USP) or off conductivity temperature compensation with 5, 10, 15, 20 or 25 °C reference temperatures.
Certifications: CE, TUV 3-in-1 and FCC Class A certifications.
Informations de livraison: Supplied with electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable, meter test certificate and 3-year replacement warranty.
The kit additionally includes: 8157BNUMD ROSS Ultra Triode refillable epoxy-body pH/ATC electrode, 013005MD DuraProbe 4-cell (K=0,475) epoxy-body conductivity sensor, ROSS pH buffer and storage solution kit and Orion 1413 µS conductivity standard (5× 60 ml).
Spécifications de la famille de produits
Spécifications
- Catalog No:
- 665-0272
- 665-0271
- Modèle:
- A215
- A215
- pH - Gamme:
- –2,000...+20,000
- –2,000...+20,000
- pH - Résolution:
- 0,1 / 0,01 / 0,001
- 0,1 / 0,01 / 0,001
- Précision - pH:
- ±0,002
- ±0,002
- pH - Étalonnage:
- Up to 5 points; Automatic buffer recognition (USA/NIST and DIN)
- Up to 5 points; Automatic buffer recognition (USA/NIST and DIN)
- mV - Gamme:
- ±2000,0
- ±2000,0
- mV - Résolution:
- 0,1
- 0,1
- mV - Précision:
- ±0,2 mV or ±0,05% of reading, whichever is greater
- ±0,2 mV or ±0,05% of reading, whichever is greater
- Plage de conductivité:
- 0,001 µS/cm - 3000 mS/cm
- 0,001 µS/cm - 3000 mS/cm
- Résolution de conductivité:
- 0,001 μS minimum, auto ranging, up to 4 significant digits
- 0,001 μS minimum, auto ranging, up to 4 significant digits
- Précision de la conductivité:
- 0,5% of reading ±1 digit > 3 μS, 0,5% of reading ±0,01 μS ≤ 3 μS
- 0,5% of reading ±1 digit > 3 μS, 0,5% of reading ±0,01 μS ≤ 3 μS
- Constante de cellule de conductivité:
- 0,001 - 199,9 cm⁻¹
- 0,001 - 199,9 cm⁻¹
- Étalonnage de conductivité:
- Up to 5 points
- Up to 5 points
- Résistivité - Plage:
- 2 Ω/cm - 100 MΩ/cm
- 2 Ω/cm - 100 MΩ/cm
- Résistivité - Précision:
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- Salinité - Gamme:
- 0,06 - 80,00 psu / 0,05 - 42,00 ppt
- 0,06 - 80,00 psu / 0,05 - 42,00 ppt
- Détermination de la résistivité:
- 1 ohm or 0,1 meg-ohm, auto ranging
- 1 ohm or 0,1 meg-ohm, auto ranging
- Salinité - Résolution:
- 0,01 psu or 0,01 ppt minimum, auto ranging
- 0,01 psu or 0,01 ppt minimum, auto ranging
- Salinité - Précision:
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- TDS - Gamme:
- 0,001 - 200,0 ppm
- 0,001 - 200,0 ppm
- TDS - Résolution:
- 0,001 ppm minimum, auto ranging, up to 4 significant digits
- 0,001 ppm minimum, auto ranging, up to 4 significant digits
- TDS - Précision:
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- Gamme de température:
- –5,0...+105,0 °C
- –5,0...+105,0 °C
- Détermination dela température:
- 0,1 °C
- 0,1 °C
- Exactitude de la températue:
- ±0,1 °C
- ±0,1 °C
- Compensation de température:
- Linear, non linear nLFn, non linear nLFu, EP (USP), off
- Linear, non linear nLFn, non linear nLFu, EP (USP), off
- Type de sonde:
- DuraProbe 4-cell conductivity sensor
- DuraProbe 4-cell conductivity sensor
- Affichage:
- Graphic LCD with backlight
- Graphic LCD with backlight
- Batterie - Durée de vie:
- 800 h
- 800 h
- Alimentation:
- AC adapter or 4 AA batteries
- AC adapter or 4 AA batteries
- Conformités:
- IP 54
- IP 54
- Température de service:
- 5...45 °C (ambient)
- 5...45 °C (ambient)
- Acquisition des données - Enregistrement:
- Manual / Automatic with AUTO-READ™
- Manual / Automatic with AUTO-READ™
- Acquisition des données - Capacité:
- 2000 with date and time stamp
- 2000 with date and time stamp
- Indice IP:
- IP 54
- IP 54
- Connexions:
- BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)
- BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)
- Sorties:
- RS232, USB
- RS232, USB
- Entrée:
- BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)
- BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)
- Température de référence:
- 5 / 10 / 15 / 20 / 25 °C
- 5 / 10 / 15 / 20 / 25 °C
- Alarme:
- High/low limit alarm, calibration due alarm
- High/low limit alarm, calibration due alarm
- Nombre de canaux:
- 2
- 2
- Étalonnage:
- pH with calibration editing option, relative mV (RmV), ORP, conductivity with calibration editing option
- pH with calibration editing option, relative mV (RmV), ORP, conductivity with calibration editing option
- Points d’étalonnage:
- 1 point relative mV (RmV) mode, 1 point ORP mode
- 1 point relative mV (RmV) mode, 1 point ORP mode
- L×P×H:
- 240×180×110 mm
- 240×180×110 mm
- Poids:
- 0,9 kg
- 0,9 kg
Spécifications
Product Family Options
Informations sur le produit
- DésignationDisponibilitéPrix
- Bench pH/conductivity meter Orion Star A215, kit incl. ROSS Ultra Triode pH/ATC probe and DuraProbe conductivity cellConnexion Pour la disponibilité1 * 1 STSFr. 2’870.00Spécifications :
Plus
Specifications
Cat. No.665-0272ModèleA215pH - Gamme–2,000...+20,000pH - Résolution0,1 / 0,01 / 0,001Précision - pH±0,002pH - ÉtalonnageUp to 5 points; Automatic buffer recognition (USA/NIST and DIN)mV - Gamme±2000,0mV - Résolution0,1mV - Précision±0,2 mV or ±0,05% of reading, whichever is greaterPlage de conductivité0,001 µS/cm - 3000 mS/cmRésolution de conductivité0,001 μS minimum, auto ranging, up to 4 significant digitsPrécision de la conductivité0,5% of reading ±1 digit > 3 μS, 0,5% of reading ±0,01 μS ≤ 3 μSConstante de cellule de conductivité0,001 - 199,9 cm⁻¹Étalonnage de conductivitéUp to 5 pointsRésistivité - Plage2 Ω/cm - 100 MΩ/cmRésistivité - Précision0,5% of reading ±1 digitSalinité - Gamme0,06 - 80,00 psu / 0,05 - 42,00 pptDétermination de la résistivité1 ohm or 0,1 meg-ohm, auto rangingSalinité - Résolution0,01 psu or 0,01 ppt minimum, auto rangingSalinité - Précision0,5% of reading ±1 digitTDS - Gamme0,001 - 200,0 ppmTDS - Résolution0,001 ppm minimum, auto ranging, up to 4 significant digitsTDS - Précision0,5% of reading ±1 digitGamme de température–5,0...+105,0 °CDétermination dela température0,1 °CExactitude de la températue±0,1 °CCompensation de températureLinear, non linear nLFn, non linear nLFu, EP (USP), offType de sondeDuraProbe 4-cell conductivity sensorAffichageGraphic LCD with backlightBatterie - Durée de vie800 hAlimentationAC adapter or 4 AA batteriesConformitésIP 54Température de service5...45 °C (ambient)Acquisition des données - EnregistrementManual / Automatic with AUTO-READ™Acquisition des données - Capacité2000 with date and time stampIndice IPIP 54ConnexionsBNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)SortiesRS232, USBEntréeBNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)Température de référence5 / 10 / 15 / 20 / 25 °CAlarmeHigh/low limit alarm, calibration due alarmNombre de canaux2ÉtalonnagepH with calibration editing option, relative mV (RmV), ORP, conductivity with calibration editing optionPoints d’étalonnage1 point relative mV (RmV) mode, 1 point ORP modeL×P×H240×180×110 mmPoids0,9 kg - Bench pH/conductivity meter Orion Star A215, instrument aloneConnexion Pour la disponibilité1 * 1 STSFr. 1’830.00Spécifications :
Plus
Specifications
Cat. No.665-0271ModèleA215pH - Gamme–2,000...+20,000pH - Résolution0,1 / 0,01 / 0,001Précision - pH±0,002pH - ÉtalonnageUp to 5 points; Automatic buffer recognition (USA/NIST and DIN)mV - Gamme±2000,0mV - Résolution0,1mV - Précision±0,2 mV or ±0,05% of reading, whichever is greaterPlage de conductivité0,001 µS/cm - 3000 mS/cmRésolution de conductivité0,001 μS minimum, auto ranging, up to 4 significant digitsPrécision de la conductivité0,5% of reading ±1 digit > 3 μS, 0,5% of reading ±0,01 μS ≤ 3 μSConstante de cellule de conductivité0,001 - 199,9 cm⁻¹Étalonnage de conductivitéUp to 5 pointsRésistivité - Plage2 Ω/cm - 100 MΩ/cmRésistivité - Précision0,5% of reading ±1 digitSalinité - Gamme0,06 - 80,00 psu / 0,05 - 42,00 pptDétermination de la résistivité1 ohm or 0,1 meg-ohm, auto rangingSalinité - Résolution0,01 psu or 0,01 ppt minimum, auto rangingSalinité - Précision0,5% of reading ±1 digitTDS - Gamme0,001 - 200,0 ppmTDS - Résolution0,001 ppm minimum, auto ranging, up to 4 significant digitsTDS - Précision0,5% of reading ±1 digitGamme de température–5,0...+105,0 °CDétermination dela température0,1 °CExactitude de la températue±0,1 °CCompensation de températureLinear, non linear nLFn, non linear nLFu, EP (USP), offType de sondeDuraProbe 4-cell conductivity sensorAffichageGraphic LCD with backlightBatterie - Durée de vie800 hAlimentationAC adapter or 4 AA batteriesConformitésIP 54Température de service5...45 °C (ambient)Acquisition des données - EnregistrementManual / Automatic with AUTO-READ™Acquisition des données - Capacité2000 with date and time stampIndice IPIP 54ConnexionsBNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)SortiesRS232, USBEntréeBNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)Température de référence5 / 10 / 15 / 20 / 25 °CAlarmeHigh/low limit alarm, calibration due alarmNombre de canaux2ÉtalonnagepH with calibration editing option, relative mV (RmV), ORP, conductivity with calibration editing optionPoints d’étalonnage1 point relative mV (RmV) mode, 1 point ORP modeL×P×H240×180×110 mmPoids0,9 kg
Accessories
Informations sur le produit
- N° de catalogueDésignationPourDisponibilitéPrix
- 662-1140Printer and computer cable set, incl. one RS232 printer interface cable and one RS232 computer interface cableThermo Scientific™ Orion™ Versa Star meters, Orion™ Star A series meters (Star A320, Star A220 and Star A210 series) and legacy Orion™ Star and Star Plus metersConnexion Pour la disponibilité1 * 1 SET - 94.40 SFr.
Informations sur le produit
- N° de catalogueDésignationDisponibilitéPrix
- 662-1547Adaptateur universel supplémentaire de rechange pour appareils de mesure de paillasse et portables Thermo Scientific™ Orion™ Star A et appareils de mesure hérités Orion™ Star et Star PlusConnexion Pour la disponibilité1 * 1 ST - 89.50 SFr.
Informations sur le produit
- N° de catalogueDésignationDisponibilitéPrix
- ORIOSTARA-HBBase lestée Orion™ Star™ A pour utilisation avec un support d'électrode relié à l'appareil de mesureConnexion Pour la disponibilité1 * 1 ST - 69.60 SFr.
Informations sur le produit
- N° de catalogueDésignationPourDisponibilitéPrix
- ORIO1010005Adaptateur de câble, pour raccorder le port RS232 d'un appareil de mesure à un port USBAppareils de mesure Orion™ Versa Star, Orion™ Star A (A320, A220 + A210) et Orion™ Star et Star PlusConnexion Pour la disponibilité1 * 1 ST - 162.00 SFr.
Informations sur le produit
- N° de catalogueDésignationPourDisponibilitéPrix
- ORIO1010053Câble RS232, pour raccorder un appareil de mesure à un PC ou une imprimanteAppareils de mesure de paillasse Orion™ Versa Star, Orion™ Star A (A320, A220 + A210), Orion™ Dual Star, et appareils de mesure Orion 3-Star, 4-Star et 5-StarConnexion Pour la disponibilité1 * 1 ST - 56.10 SFr.
Informations sur le produit
- N° de catalogueDésignationDisponibilitéPrix
- ORIOSTARA-BEASupport d'électrode relié à un appareil de mesure pour paillasse (bras et tête de support d'électrode), Orion™ Star A, mouvement à 360° (sans base lestée)Connexion Pour la disponibilité1 * 1 ST - 202.00 SFr.
Les recommandations seront personnalisées selon vos préférences d’achat uniquement si vous avez activé "Améliorer mon expérience d’achat" dans la page "Informations personnelles".
Sinon, vous recevrez des recommandations génériques.