Order Entry
Switzerland
ContactUsLinkComponent
 
Temperature loggers, ECOLOG TP2
  ELPR800446
undefined
Temperature loggers, ECOLOG TP2
  ELPR800446

 

  • Plage de mesure:
    T: –35…+55 °C
  • Data memory - No. of readings:
    64000
  • Interface:
    RS232
  • Désignation:
    Temperature logger, ECOLOG TP2, two channel for EX zone 1
  • Seq:
    02
  • Modèle:
    ECOLOG TP2-Ex
  • Batterie - Durée de vie:
    1,5 years
  • Alarme:
    External on DB15, and alarm display on LCD screen (programmable)
  • Piles:
    One lithium battery, 3,6 V
  • Affichage:
    LCD
  • Intervalles d’enregistrement:
    1 min
  • Sonde:
    2×Pt100 sensors
  • L×P×H:
    110×85×35 mm
  • Précision:
    ±0,3 °C (–99,9…500 °C);
    ±0,5 °C (500,1…550 °C);
    ±0,6 °C (–200...–100 °C)
  • Cdt:
    1

 

 

ECOLOG TP2 is a two channel data loggers for the measurement of temperature. These loggers suitable for foodstuff applications. It has desired time interval (from every 1 second to every 3 hours).

  • Low battery warning
  • Direct connection to printer for alarm protocol and status
  • High and low resolution selectable
  • Pt100 sensors of –200 to +550 °C with four wire system

Having thermoplastic ABS mount, keypad have four keys: reset alarm, step by step data or alarm display, printout data/alarm.

PC software elproLOG ANALYZE for all communication, reprogramming, display, statistics and printout.

 : IP 52 rated protection class.

Probes