Conductivity meter, handheld, Eutech COND 6+
Fournisseur: EUTECH
Certificats
À propos de cet article
This portable economic meter with large LCD display is ideal for quick, accurate conductivity checks in laboratories, field and schools.
- Option of quick, easy automatic calibration or customised, near-to-sample manual calibration
- Auto-ranging for conductivity measurements for fast response and best resolution over wide measurement range
- HOLD function and auto-power off
- Self-diagnostic with message codes for easy troubleshooting
- Electrode with built-in ATC designed for minimal air bubble entrapment during measurement
- Built with shock-resistant ABS plastic and protected by an IP54-rated housing against accidental splashes
- Protective rubber boot shields meter from drops and features a sturdy built-in stand for easy bench top operation
Useful in nutrient and fertiliser checks in hydroponics and agricultural industries. Also suitable for analysing water hard water, untreated water, industrial and rinse water, drinking water, effluent water, pool water and incoming process water. Ideal for all types of quality assurance, printing industries and water quality testing.
Livraison: Supplied with conductivity electrode ECCONSEN91B.
The kit additionally includes the conductivity carrying kit (includes 1× 1413 uS, 12,88 mS, 84 uS KCl standard solution and 1× deionised (rinse) water (60 ml each).
The kit additionally includes the conductivity carrying kit (includes 1× 1413 uS, 12,88 mS, 84 uS KCl standard solution and 1× deionised (rinse) water (60 ml each).
Spécifications
- Catalog No:
- 663-0353
- 663-0350
- Modèle:
- Eutech COND 6+
- Eutech COND 6+
- Plage de conductivité:
- 0 - 19,99 uS/cm
19,9 - 199,9 uS/cm
199,9 - 1999 µS/cm
2,00 - 19,99 mS/cm
20,0 - 199,9 mS/cm - 0 - 19,99 uS/cm
19,9 - 199,9 uS/cm
199,9 - 1999 µS/cm
2,00 - 19,99 mS/cm
20,0 - 199,9 mS/cm
- Résolution de conductivité:
- 0,05 % full scale
- 0,05 % full scale
- Précision de la conductivité:
- 1% full scale + 1 LSD
- 1% full scale + 1 LSD
- Constante de cellule de conductivité:
- 0,1 / 1,0 / 10,0
- 0,1 / 1,0 / 10,0
- Étalonnage de conductivité:
- 4 points (1 per range) auto, 5 points (1 per range) manual
- 4 points (1 per range) auto, 5 points (1 per range) manual
- Gamme de température:
- 0,0...+80,0 °C
- 0,0...+80,0 °C
- Détermination dela température:
- 0,1 °C
- 0,1 °C
- Exactitude de la températue:
- ±0,5 °C
- ±0,5 °C
- Compensation de température:
- ATC/MTC (0 - 80 °C) (with supplied probe)
- ATC/MTC (0 - 80 °C) (with supplied probe)
- Coefficient de température:
- 0,0 - 3,0 %
- 0,0 - 3,0 %
- Connexions:
- BNC, 2,5 mm phono socket
- BNC, 2,5 mm phono socket
- Alimentation:
- 4× 1,5 V ‘AAA’ alkaline batteries
- 4× 1,5 V ‘AAA’ alkaline batteries
- Affichage:
- LCD
- LCD
- Conformités:
- IP54
- IP54
- L×P×H:
- 85×42×157 cm
- 85×42×157 cm
- Poids:
- 0,255 kg
- 0,255 kg
Spécifications