pH/mV/°C meter, handheld, Orion Star™ A321
 
undefined
pH/mV/°C meter, handheld, Orion Star™ A321

 

  • Catalog No:
  • 662-1446
  • 662-1447
  • Modèle:
  • A321
  • A321
  • pH - Gamme:
  • −2,000...+20,000
  • −2,000...+20,000
  • pH - Résolution:
  • 0,1 / 0,01 / 0,001
  • 0,1 / 0,01 / 0,001
  • Précision - pH:
  • ±0,002
  • ±0,002
  • pH - Étalonnage:
  • Up to 5 points; Automatic buffer recognition (USA/NIST and DIN)
  • Up to 5 points; Automatic buffer recognition (USA/NIST and DIN)
  • mV - Gamme:
  • ±2000,0
  • ±2000,0
  • mV - Résolution:
  • 0,1
  • 0,1
  • mV - Précision:
  • ±0,2 mV or ±0,05% of reading, whichever is greater
  • ±0,2 mV or ±0,05% of reading, whichever is greater
  • Gamme de température:
  • −5,0...+105,0 °C
  • −5,0...+105,0 °C
  • Détermination dela température:
  • 0,1 °C
  • 0,1 °C
  • Exactitude de la températue:
  • ±0,1 °C
  • ±0,1 °C
  • Compensation de température:
  • Automatic or manual
  • Automatic or manual
  • Affichage:
  • Graphic LCD with backlight
  • Graphic LCD with backlight
  • Batterie - Durée de vie:
  • 800 h.
  • 800 h.
  • Alimentation:
  • AC adapter (not included) or 4 AA batteries (included)
  • AC adapter (not included) or 4 AA batteries (included)
  • Conformités:
  • IP 67
  • IP 67
  • Température de service:
  • 5...45 °C (ambient)
  • 5...45 °C (ambient)
  • Acquisition des données - Enregistrement:
  • Manual / Automatic with AUTO-READ™, timed
  • Manual / Automatic with AUTO-READ™, timed
  • Acquisition des données - Capacité:
  • 5000 with date and time stamp
  • 5000 with date and time stamp
  • Mémoire:
  • Non-volatile memory preserves data log, calibration log and meter settings
  • Non-volatile memory preserves data log, calibration log and meter settings
  • Langues:
  • English, Spanish, French, Italian, German and Chinese; Portuguese and Korean available with software update on web
  • English, Spanish, French, Italian, German and Chinese; Portuguese and Korean available with software update on web
  • Connexions:
  • BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (ATC temperature probe)
  • BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (ATC temperature probe)
  • Sorties:
  • RS232, USB
  • RS232, USB
  • Alarme:
  • High/low limit, calibration due alarm
  • High/low limit, calibration due alarm
  • Nombre de canaux:
  • 1
  • 1
  • Étalonnage:
  • pH with calibration editing option, relative mV (RmV), ORP and temperature
  • pH with calibration editing option, relative mV (RmV), ORP and temperature
  • Points d’étalonnage:
  • 1 - 5
  • 1 - 5
  • L×P×H:
  • 240×105×60 mm
  • 240×105×60 mm
  • Poids:
  • 0,5 kg
  • 0,5 kg

 