Saisie de commande
Belgium
Lien pour commandesContactUsLinkComponent
Orion Star™ A329 Multi-Parameter Meter (pH/ISE/Conductivity/DO), Handheld

Orion Star™ A329 Multi-Parameter Meter (pH/ISE/Conductivity/DO), Handheld

Supplier:   Thermo Orion
Orion Star™ A329 Multi-Parameter Meter (pH/ISE/Conductivity/DO), Handheld
Supplier:  Thermo Orion
Certificates
Orion Star™ A329 Multi-Parameter Meter (pH/ISE/Conductivity/DO), Handheld
Orion Star™ A329 Multi-Parameter Meter (pH/ISE/Conductivity/DO), Handheld

2 Options Available Below

Product Details & Documents

This rugged, waterproof portable meter is designed for a wide range of pH, ion concentration, conductivity, dissolved oxygen and temperature testing and field applications. It can be used in the most demanding locations thanks to its IP67-rated housing. The meter offers three measuring channels allowing users to measure pH, conductivity and DO simultaneously or view each channel separately.

  • Up to five point pH calibrations with calibration editing option that allows individual points to be fixed without a full recalibration
  • Automatically compensate dissolved oxygen readings using ATC temperature, manual salinity input and built-in barometer for barometric pressure readings
  • Informative, easy-to-read backlit graphic display with clear onscreen instructions simplifies operation
  • Measurement stability indicator and selectable read modes – Auto-Read holds a stable reading, Continuous shows changing readings with hold option and Timed logs data at specific intervals
  • Data log collects and date/time stamps up to 5000 measurement sets with optional sample ID and user ID
  • Calibration log stores ten most recent calibrations
  • Orion Star Com software facilitates data transfer to a computer via USB or RS232 connections

Measure pH, mV, relative mV or ORP with temperature on channel one and measure dissolved oxygen in percent saturation or mg/L with temperature using RDO optical DO probes or polarographic DO probes on channel two.

Automatically compensate dissolved oxygen readings using ATC temperature, manual salinity input and built-in barometer for barometric pressure readings. Transfer single, range or all data points to printer or computer.

Auto-ranging conductivity values offer consistent readings across entire measurement range without need to recalibrate. Use linear, non linear, EP (USP) or off conductivity temperature compensation with 5, 10, 15, 20 or 25 °C reference temperatures. 3-years replacement warranty.

Certifications: TUV 3-1, FCC Class A.

Delivery information: Supplied with four with 4 AA batteries, literature CD, printed quick start guide, computer interface cable and meter test certificate.
The kit additionally includes: 8107UWMMD ROSS Ultra Triode gel filled epoxy-body pH/ATC electrode (3 m cable), 013010MD DuraProbe 4-cell (K=0,475) epoxy-body conductivity sensor (3 m cable), 087010MD RDO optical DO sensor (3 m cable), calibration sleeve and guard, Orion pH 4/7/10 buffer pouches (10 packs each), ROSS electrode storage solution (475 ml), Orion rinse solution pouches (10 packs), Orion 1413 µS conductivity standard pouches (10 packs), Orion Star A series portable meter armor and Orion Star A series hard-sided field case.

Specifications for Product Family

Specifications

  • Catalog No:
  • 665-0281
  • 665-0282
  • Modèle:
  • A329
  • A329
  • pH - Gamme:
  • –2,000...+20,000
  • –2,000...+20,000
  • pH - Résolution:
  • 0,1 / 0,01 / 0,001
  • 0,1 / 0,01 / 0,001
  • Précision - pH:
  • ±0,002
  • ±0,002
  • pH - Étalonnage:
  • Up to 5 points; Automatic buffer recognition (USA/NIST and DIN)
  • Up to 5 points; Automatic buffer recognition (USA/NIST and DIN)
  • mV - Gamme:
  • ±2000,0
  • ±2000,0
  • mV - Résolution:
  • 0,1
  • 0,1
  • mV - Précision:
  • ±0,2 mV or 0,05%
  • ±0,2 mV or 0,05%
  • ISE Gamme:
  • 0,0001 - 19900
  • 0,0001 - 19900
  • ISE Résolution:
  • 0,0001 minimum, 1 to 3 significant digits (user selectable)
  • 0,0001 minimum, 1 to 3 significant digits (user selectable)
  • ISE Précision:
  • ±0,2 mV or ±0,05% of reading, whichever is greater
  • ±0,2 mV or ±0,05% of reading, whichever is greater
  • ISE Étalonnage:
  • Up to 5 points
  • Up to 5 points
  • ISE Unités:
  • ppm, M, mg/l, %, ppb, none
  • ppm, M, mg/l, %, ppb, none
  • Plage de conductivité:
  • 0,001 µS/cm - 3000 mS/cm
  • 0,001 µS/cm - 3000 mS/cm
  • Résolution de conductivité:
  • 0,001 μS minimum, auto ranging, up to 4 significant digits
  • 0,001 μS minimum, auto ranging, up to 4 significant digits
  • Précision de la conductivité:
  • ±0,5% of reading ±1 digit
  • ±0,5% of reading ±1 digit
  • Constante de cellule de conductivité:
  • 0,001 - 199,9 cm⁻¹
  • 0,001 - 199,9 cm⁻¹
  • Étalonnage de conductivité:
  • Up to 5 points
  • Up to 5 points
  • Température de référence de conductivité:
  • 5/10/15/20/25 °C
  • 5/10/15/20/25 °C
  • Résistivité - Plage:
  • 2 Ω/cm - 100 MΩ/cm
  • 2 Ω/cm - 100 MΩ/cm
  • Résistivité - Précision:
  • 0,5% of reading ±1 digit
  • 0,5% of reading ±1 digit
  • Salinité - Gamme:
  • 0,06 - 80,00 psu / 0,05 - 42,00 ppt
  • 0,06 - 80,00 psu / 0,05 - 42,00 ppt
  • Détermination de la résistivité:
  • 1 ohm or 0,1 meg-ohm, auto ranging
  • 1 ohm or 0,1 meg-ohm, auto ranging
  • Salinité - Résolution:
  • 0,01 psu or 0,01 ppt minimum, auto ranging
  • 0,01 psu or 0,01 ppt minimum, auto ranging
  • Salinité - Précision:
  • 0,5% of reading ±1 digit
  • 0,5% of reading ±1 digit
  • TDS - Gamme:
  • 0,001 - 200,0 ppm
  • 0,001 - 200,0 ppm
  • TDS - Résolution:
  • 0,001 ppm minimum, auto ranging, up to 4 significant digits
  • 0,001 ppm minimum, auto ranging, up to 4 significant digits
  • TDS - Précision:
  • 0,5% of reading ±1 digit
  • 0,5% of reading ±1 digit
  • TDS - Facteur de conversion:
  • Linear (0,02 - 9,99)
  • Linear (0,02 - 9,99)
  • OD Concentration:
  • 0 - 90 mg/L (polarographic); 0 - 50 mg/L (RDO)
  • 0 - 90 mg/L (polarographic); 0 - 50 mg/L (RDO)
  • Oxygène dissous - Résolution:
  • 0,1 / 0,01 mg/L
  • 0,1 / 0,01 mg/L
  • Oxygène dissous - saturation:
  • 0 - 600 % (polarographic); 0 - 500 % (RDO)
  • 0 - 600 % (polarographic); 0 - 500 % (RDO)
  • OD Précision:
  • Polarographic: ±0,2 mg/l or ±2% of reading, whichever is greater
    RDO: ±0,1 mg/l (up to 8 mg/), ±0,2 mg/l (8 - 20 mg/l), ±10% of reading (up to 50 mg/l)
  • Polarographic: ±0,2 mg/l or ±2% of reading, whichever is greater
    RDO: ±0,1 mg/l (up to 8 mg/), ±0,2 mg/l (8 - 20 mg/l), ±10% of reading (up to 50 mg/l)
  • OD Etalonnage:
  • Water-saturated air, air saturated water, manual (Winkler) and zero point
  • Water-saturated air, air saturated water, manual (Winkler) and zero point
  • Oxygène dissous - Gamme de pression barométrique:
  • Automatic or manual, 450 - 850 mmHg
  • Automatic or manual, 450 - 850 mmHg
  • Gamme de température:
  • –5,0...+105,0 °C (–5,0...+50,0 °C with DO)
  • –5,0...+105,0 °C (–5,0...+50,0 °C with DO)
  • Détermination dela température:
  • 0,1 °C
  • 0,1 °C
  • Exactitude de la températue:
  • ±0,1 °C
  • ±0,1 °C
  • Calibrage de température:
  • 1 point temperature offset calibration
  • 1 point temperature offset calibration
  • Compensation de température:
  • Automatic or manual
  • Automatic or manual
  • Gamme de pression:
  • 400 - 850 mm Hg
  • 400 - 850 mm Hg
  • Détermination de la pression:
  • 1 mm Hg
  • 1 mm Hg
  • Affichage:
  • Graphic LCD with backlight
  • Graphic LCD with backlight
  • Type de batterie:
  • 4 AA (included)
  • 4 AA (included)
  • Batterie - Durée de vie:
  • 800 hours
  • 800 hours
  • Alimentation:
  • AC adapter or 4 AA batteries
  • AC adapter or 4 AA batteries
  • Conformités:
  • IP 67
  • IP 67
  • Acquisition des données - Enregistrement:
  • Manual / Automatic with AUTO-READ™
  • Manual / Automatic with AUTO-READ™
  • Acquisition des données - Capacité:
  • 5000
  • 5000
  • Mémoire:
  • Non-volatile memory preserves data log, calibration log and meter settings
  • Non-volatile memory preserves data log, calibration log and meter settings
  • Langues:
  • English, Spanish, French, Italian, German and Chinese; Portuguese and Korean available with software update
  • English, Spanish, French, Italian, German and Chinese; Portuguese and Korean available with software update
  • Indice IP:
  • IP 67
  • IP 67
  • Connexions:
  • BNC, pin tip, 8- and 9-pin mini-DIN
  • BNC, pin tip, 8- and 9-pin mini-DIN
  • Sorties:
  • RS232, USB
  • RS232, USB
  • Entrée:
  • 3 measuring channels
  • 3 measuring channels
  • Alarme:
  • High/low limit alarm, calibration due alarm
  • High/low limit alarm, calibration due alarm
  • Nombre de canaux:
  • 3
  • 3
  • Étalonnage:
  • pH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ
  • pH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ
  • Points d’étalonnage:
  • 1 point relative mV (RmV) mode, 1 point ORP mode
  • 1 point relative mV (RmV) mode, 1 point ORP mode
  • L×P×H:
  • 60×105×240 mm
  • 60×105×240 mm
  • Poids:
  • 0,5 kg
  • 0,5 kg

Specifications

Product Family Options

Product Information

  • Désignation
    Availability
    Price
  • Handheld pH/ISE/conductivity/DO meter Orion Star A329, instrument alone
    Log in for availability
    Log in for availability
    1 * 1 ST
    1 * 1 ST - 2 520,00 €
    2 520,00 €
    1 * 1 ST
       2 520,00 €

    Specifications:
    More
    Specifications
    Cat. No.
    665-0281
    Modèle
    A329
    pH - Gamme
    –2,000...+20,000
    pH - Résolution
    0,1 / 0,01 / 0,001
    Précision - pH
    ±0,002
    pH - Étalonnage
    Up to 5 points; Automatic buffer recognition (USA/NIST and DIN)
    mV - Gamme
    ±2000,0
    mV - Résolution
    0,1
    mV - Précision
    ±0,2 mV or 0,05%
    ISE Gamme
    0,0001 - 19900
    ISE Résolution
    0,0001 minimum, 1 to 3 significant digits (user selectable)
    ISE Précision
    ±0,2 mV or ±0,05% of reading, whichever is greater
    ISE Étalonnage
    Up to 5 points
    ISE Unités
    ppm, M, mg/l, %, ppb, none
    Plage de conductivité
    0,001 µS/cm - 3000 mS/cm
    Résolution de conductivité
    0,001 μS minimum, auto ranging, up to 4 significant digits
    Précision de la conductivité
    ±0,5% of reading ±1 digit
    Constante de cellule de conductivité
    0,001 - 199,9 cm⁻¹
    Étalonnage de conductivité
    Up to 5 points
    Température de référence de conductivité
    5/10/15/20/25 °C
    Résistivité - Plage
    2 Ω/cm - 100 MΩ/cm
    Résistivité - Précision
    0,5% of reading ±1 digit
    Salinité - Gamme
    0,06 - 80,00 psu / 0,05 - 42,00 ppt
    Détermination de la résistivité
    1 ohm or 0,1 meg-ohm, auto ranging
    Salinité - Résolution
    0,01 psu or 0,01 ppt minimum, auto ranging
    Salinité - Précision
    0,5% of reading ±1 digit
    TDS - Gamme
    0,001 - 200,0 ppm
    TDS - Résolution
    0,001 ppm minimum, auto ranging, up to 4 significant digits
    TDS - Précision
    0,5% of reading ±1 digit
    TDS - Facteur de conversion
    Linear (0,02 - 9,99)
    OD Concentration
    0 - 90 mg/L (polarographic); 0 - 50 mg/L (RDO)
    Oxygène dissous - Résolution
    0,1 / 0,01 mg/L
    Oxygène dissous - saturation
    0 - 600 % (polarographic); 0 - 500 % (RDO)
    OD Précision
    Polarographic: ±0,2 mg/l or ±2% of reading, whichever is greater
    RDO: ±0,1 mg/l (up to 8 mg/), ±0,2 mg/l (8 - 20 mg/l), ±10% of reading (up to 50 mg/l)
    OD Etalonnage
    Water-saturated air, air saturated water, manual (Winkler) and zero point
    Oxygène dissous - Gamme de pression barométrique
    Automatic or manual, 450 - 850 mmHg
    Gamme de température
    –5,0...+105,0 °C (–5,0...+50,0 °C with DO)
    Détermination dela température
    0,1 °C
    Exactitude de la températue
    ±0,1 °C
    Calibrage de température
    1 point temperature offset calibration
    Compensation de température
    Automatic or manual
    Gamme de pression
    400 - 850 mm Hg
    Détermination de la pression
    1 mm Hg
    Affichage
    Graphic LCD with backlight
    Type de batterie
    4 AA (included)
    Batterie - Durée de vie
    800 hours
    Alimentation
    AC adapter or 4 AA batteries
    Conformités
    IP 67
    Acquisition des données - Enregistrement
    Manual / Automatic with AUTO-READ™
    Acquisition des données - Capacité
    5000
    Mémoire
    Non-volatile memory preserves data log, calibration log and meter settings
    Langues
    English, Spanish, French, Italian, German and Chinese; Portuguese and Korean available with software update
    Indice IP
    IP 67
    Connexions
    BNC, pin tip, 8- and 9-pin mini-DIN
    Sorties
    RS232, USB
    Entrée
    3 measuring channels
    Alarme
    High/low limit alarm, calibration due alarm
    Nombre de canaux
    3
    Étalonnage
    pH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ
    Points d’étalonnage
    1 point relative mV (RmV) mode, 1 point ORP mode
    L×P×H
    60×105×240 mm
    Poids
    0,5 kg
  • Handheld pH/ISE/conductivity/DO meter Orion Star A329, kit incl. pH electrode, conductivity cell and RDO optical sensor
    Log in for availability
    Log in for availability
    1 * 1 ST
    1 * 1 ST - 3 800,00 €
    3 800,00 €
    1 * 1 ST
       3 800,00 €

    Specifications:
    More
    Specifications
    Cat. No.
    665-0282
    Modèle
    A329
    pH - Gamme
    –2,000...+20,000
    pH - Résolution
    0,1 / 0,01 / 0,001
    Précision - pH
    ±0,002
    pH - Étalonnage
    Up to 5 points; Automatic buffer recognition (USA/NIST and DIN)
    mV - Gamme
    ±2000,0
    mV - Résolution
    0,1
    mV - Précision
    ±0,2 mV or 0,05%
    ISE Gamme
    0,0001 - 19900
    ISE Résolution
    0,0001 minimum, 1 to 3 significant digits (user selectable)
    ISE Précision
    ±0,2 mV or ±0,05% of reading, whichever is greater
    ISE Étalonnage
    Up to 5 points
    ISE Unités
    ppm, M, mg/l, %, ppb, none
    Plage de conductivité
    0,001 µS/cm - 3000 mS/cm
    Résolution de conductivité
    0,001 μS minimum, auto ranging, up to 4 significant digits
    Précision de la conductivité
    ±0,5% of reading ±1 digit
    Constante de cellule de conductivité
    0,001 - 199,9 cm⁻¹
    Étalonnage de conductivité
    Up to 5 points
    Température de référence de conductivité
    5/10/15/20/25 °C
    Résistivité - Plage
    2 Ω/cm - 100 MΩ/cm
    Résistivité - Précision
    0,5% of reading ±1 digit
    Salinité - Gamme
    0,06 - 80,00 psu / 0,05 - 42,00 ppt
    Détermination de la résistivité
    1 ohm or 0,1 meg-ohm, auto ranging
    Salinité - Résolution
    0,01 psu or 0,01 ppt minimum, auto ranging
    Salinité - Précision
    0,5% of reading ±1 digit
    TDS - Gamme
    0,001 - 200,0 ppm
    TDS - Résolution
    0,001 ppm minimum, auto ranging, up to 4 significant digits
    TDS - Précision
    0,5% of reading ±1 digit
    TDS - Facteur de conversion
    Linear (0,02 - 9,99)
    OD Concentration
    0 - 90 mg/L (polarographic); 0 - 50 mg/L (RDO)
    Oxygène dissous - Résolution
    0,1 / 0,01 mg/L
    Oxygène dissous - saturation
    0 - 600 % (polarographic); 0 - 500 % (RDO)
    OD Précision
    Polarographic: ±0,2 mg/l or ±2% of reading, whichever is greater
    RDO: ±0,1 mg/l (up to 8 mg/), ±0,2 mg/l (8 - 20 mg/l), ±10% of reading (up to 50 mg/l)
    OD Etalonnage
    Water-saturated air, air saturated water, manual (Winkler) and zero point
    Oxygène dissous - Gamme de pression barométrique
    Automatic or manual, 450 - 850 mmHg
    Gamme de température
    –5,0...+105,0 °C (–5,0...+50,0 °C with DO)
    Détermination dela température
    0,1 °C
    Exactitude de la températue
    ±0,1 °C
    Calibrage de température
    1 point temperature offset calibration
    Compensation de température
    Automatic or manual
    Gamme de pression
    400 - 850 mm Hg
    Détermination de la pression
    1 mm Hg
    Affichage
    Graphic LCD with backlight
    Type de batterie
    4 AA (included)
    Batterie - Durée de vie
    800 hours
    Alimentation
    AC adapter or 4 AA batteries
    Conformités
    IP 67
    Acquisition des données - Enregistrement
    Manual / Automatic with AUTO-READ™
    Acquisition des données - Capacité
    5000
    Mémoire
    Non-volatile memory preserves data log, calibration log and meter settings
    Langues
    English, Spanish, French, Italian, German and Chinese; Portuguese and Korean available with software update
    Indice IP
    IP 67
    Connexions
    BNC, pin tip, 8- and 9-pin mini-DIN
    Sorties
    RS232, USB
    Entrée
    3 measuring channels
    Alarme
    High/low limit alarm, calibration due alarm
    Nombre de canaux
    3
    Étalonnage
    pH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ
    Points d’étalonnage
    1 point relative mV (RmV) mode, 1 point ORP mode
    L×P×H
    60×105×240 mm
    Poids
    0,5 kg
Accessories

Product Information

  • Catalog #
    Désignation
    Pour
    Availability
    Price
  • 662-1140
    Printer and computer cable set, incl. one RS232 printer interface cable and one RS232 computer interface cable
    Thermo Scientific™ Orion™ Versa Star meters, Orion™ Star A series meters (Star A320, Star A220 and Star A210 series) and legacy Orion™ Star and Star Plus meters
    Log in for availability
    1 * 1 SET - 94,80 €

Product Information

  • Catalog #
    Désignation
    Availability
    Price
  • ORIOSTARA-ESPH
    Orion™ Star™ A Series portable meter pH electrode holder, can be attached to the portable meter armor and holds one pH electrode when placed in the pH electrode storage sleeve
    Log in for availability
    1 * 1 ST - 36,60 €

Product Information

  • Catalog #
    Désignation
    Availability
    Price
  • ORIOSTARA-CS
    Sacoche de terrain rigide pour appareils de mesure portables, contenant un appareil de mesure (avec ou sans blindage), des supports d'électrodes portables, des piles supplémentaires, des solutions, une électrode de pH dans une gaine de stockage, une sonde de conductivité et une sonde à oxygène dissous dans une gaine de stockage
    Log in for availability
    1 * 1 ST - 179,00 €

Product Information

  • Catalog #
    Désignation
    Availability
    Price
  • 662-1547
    Adaptateur universel supplémentaire de rechange pour appareils de mesure de paillasse et portables Thermo Scientific™ Orion™ Star A et appareils de mesure hérités Orion™ Star et Star Plus
    Log in for availability
    1 * 1 ST - 89,80 €

Product Information

  • Catalog #
    Désignation
    Availability
    Price
  • ORIOSTARA-ESCD
    Conductimètre portable Orion™ Star A et support de sonde d'oxygène dissous, pouvant être fixé de n'importe quel côté du blindage de l'appareil de mesure portable, et pouvant accueillir des sondes de conductivité de 12/15 mm de Ø et une sonde à oxygène dissous lorsqu'elle est placée dans sa gaine d'étalonnage
    Log in for availability
    1 * 1 ST - 36,60 €

Product Information

  • Catalog #
    Désignation
    Availability
    Price
  • ORIOSTARA-AR
    Blindage de protection pour appareils de mesure portables, avec dragonne et support intégré, gaine de stockage d'électrode de pH, support d'électrode de pH, et support de capteur de conductivité et d'oxygène dissous
    Log in for availability
    1 * 1 ST - 127,00 €

Product Information

  • Catalog #
    Désignation
    Availability
    Price
  • ORIO1210005
    Sacoche de terrain souple, comprenant des séparateurs intérieurs et des poches extérieures qui permettent d'organiser l'appareil de mesure et les accessoires sur le terrain
    Log in for availability
    1 * 1 ST - 200,00 €

Product Information

  • Catalog #
    Désignation
    Pour
    Availability
    Price
  • ORIO1010053
    Câble RS232, pour raccorder un appareil de mesure à un PC ou une imprimante
    Appareils de mesure de paillasse Orion™ Versa Star, Orion™ Star A (A320, A220 + A210), Orion™ Dual Star, et appareils de mesure Orion 3-Star, 4-Star et 5-Star
    Log in for availability
    1 * 1 ST - 56,30 €

Product Information

  • Catalog #
    Désignation
    Pour
    Availability
    Price
  • ORIO1010005
    Adaptateur de câble, pour raccorder le port RS232 d'un appareil de mesure à un port USB
    Appareils de mesure Orion™ Versa Star, Orion™ Star A (A320, A220 + A210) et Orion™ Star et Star Plus
    Log in for availability
    1 * 1 ST - 163,00 €

Recommendations

Recommendations will be personalized based on your shopping preferences only if you have given your consent by enabling "Enhance my Shopping Experience" on the "Personal Info page".

Otherwise, you will receive generic recommendations.